Kanomax 200 Series Скачать руководство пользователя страница 17

16 

Fault Description

 

Possible cause

 

Remedy

 

Reading higher than 
expected in the pres-
ence of sensor gas

Sensor correct

Interferent gas pre-
sent

Sensor calibration lost

Check calibration of gas 
generator.

Move sensor to clean air 
and check reading upon 
exposure to known gas 
concentration

Replace /refurbish sensor

Reading unstable

Power supply unsta-
ble

Power supply current 
rating incorrect

Local air flow too high

Environmental condi-
tions fluctua ing

Install stable power supply

Install power supply with 
correct rating

Reduce air flow

Reduce fluctuations

netzerotools.com

netzerotools.com

Содержание 200 Series

Страница 1: ...1 User Guide 200 netzerotools com Kanomax S200 Handheld Multi Gas Monitor netzerotools com...

Страница 2: ...concentrations above the recommended maximum range Do not place the sensor head in a gas stream the monitor is designed to measure ambient levels of ozone and uses active sampling to achieve this 1 Se...

Страница 3: ...4 5 Zero Calibration 11 4 6 Changing Unit of Measurement 12 4 7 Max Min Average Measurement Cycle 12 4 8 Alarm Mute Perc only 13 4 9 Alarm Perc only 13 4 10 Alarm Level Perc only 14 5 0 Troubleshootin...

Страница 4: ...y of the components are missing Replacements Replacement batteries sensor heads and other accessories can be purchased separately Note All sensors deteriorate over time It will be necessary to replace...

Страница 5: ...user the ability to replace faulty sensor heads without having to replace the monitor The monitor can be mains and or battery pow ered and either permanently mounted or used as a portable unit 2 The...

Страница 6: ...be installed and charged prior to use 3 1 Battery Pack Remove the battery cover from the monitor Insert the battery power connector into the monitor power socket in the back of the monitor Lay the ba...

Страница 7: ...nitor Do not disassemble or mutilate the batteries as this may cause burns Do not incinerate or heat as this may cause burns the batter ies may burst or cause the release of toxic materials Do not sho...

Страница 8: ...the base Remember to switch off the monitor before removing the sensor head Unplugging the main power will reset the above charging condi tions The batteries can be charged and discharged many times...

Страница 9: ...rming up or sensor failure in older models to disappear Should the message remain after this time a genuine sensor failure may have occurred See the trouble shooting guide Since O3 is an oxidising gas...

Страница 10: ...the display will change to indicate the type of sensor head The monitor will warm up as described in Warm up in section 3 2 Once warm up is complete the main display will appear for example To turn t...

Страница 11: ...full power and the fan remains on 4 2 Operating under Battery power A battery indicator will appear at the bottom of the display when the monitor is operating under battery power only The indicator sh...

Страница 12: ...C 68 F 3 5 F Relative Humidity of 50 5 An incorrect ZERO CAL does not cause permanent damage The process should be repeated under the above conditions NOTE Carbon Dioxide sensors MUST be calibrated in...

Страница 13: ...s the Min Max Av routine Press to scroll to EXIT Press to Exit out of Menu Setup The following display will appear Repeat the above sequence to stop the Min Max Av routine Note The MIN MAX and AVE rea...

Страница 14: ...RM MUTE ON Press to confirm your choice and you will return to the main menu If you select Alarm Mute On and exit the menu the symbol will be displayed on the gas concentration screen Press to scroll...

Страница 15: ...een will appear Press to toggle between 50PPM 100PPM 150PPM and 35PPM Press to confirm your choice and you will return to the main menu Whatever alarm level you have chosen will be displayed on the bo...

Страница 16: ...under zero gas conditions Background gas level higher than normal Interferent gas present Sensor damaged Move sensor to clean air and recheck baseline Move sensor to clean air and recheck baseline Re...

Страница 17: ...erator Move sensor to clean air and check reading upon exposure to known gas concentration Replace refurbish sensor Reading unstable Power supply unsta ble Power supply current rating incorrect Local...

Страница 18: ...s control None Series 200 Monitor Base External signal functions None Analogue output None Power supply 12 VDC 800 mA Rechargeable battery pack 9 6V 2100mA hr Ni MH Portable Yes Permanently fixable Sc...

Страница 19: ...18 Depressions in the battery cover demarcating the screw hole posi tions for fixing the unit to the wall 7 0 Drawings netzerotools com netzerotools com...

Страница 20: ...mage the electronic componentry Do not attempt to open Non expert handling of the device may cause damage Do not drop knock or shake as this could lead to internal damage Do not use harsh chemicals cl...

Страница 21: ...hen connecting to any other device read the appropriate user guide for detailed safety instructions Do not connect in compatible products HAZARDOUS ENVIRONMENTS Ozone Sensors Only Do not use the Ozone...

Страница 22: ...protection against harmful interference in a residential installation This equipment gener ates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruc...

Страница 23: ...subsequent period as stated in the following table This warranty is expressly limited to the original owner who purchases the equip ment directly from Aeroqual or from an authorized Aeroqual dealer W...

Страница 24: ...d by use of non Aeroqual original accessories This warranty becomes void if a non Aeroqual approved AC DC adaptor or battery is used Tampering with any part of the product will void the warranty Damag...

Страница 25: ...cribed in this document without prior notice Under no circumstances shall Aeroqual be responsible for any loss of data or income or any special incidental consequential or indirect dam ages howsoever...

Страница 26: ...ler has been designed to meas ure the ambient concentration of ozone The controller must not be placed directly in an ozone stream For indoor local area monitoring attach the controller to an inert su...

Страница 27: ...d in the past week There are trace levels of gases present in the air to which the sensor is cross sensitive In many cases the non zero base line reading occurs in an environment which is supposed to...

Страница 28: ...ne standard 0 25 ppm Ozone O3 0 0 150 ppm 0 0 5 ppm 0 10 ppm Leak detector Perchloroethylene C2Cl4 0 200 ppm Sulphur dioxide SO2 0 10 ppm 0 100 ppm Volatile Organic Compounds VOC Isobutylene standard...

Отзывы: