KANGAROO E-PUMP INSTRUCTIONS FOR FEED/FLUSH ONLY
(adapted from Kangaroo ePump Instructions for Use Manual)
CC.12.10A
BC Children’s Hospital Child & Youth Health Policy and Procedure Manual
Page 2 of 2
May 1
st
, 2012
Refer to online version – Print copy may not be current – Discard after use
o
Press
RUN
to restart pump at anytime
8. To clear feed totals (“
mL FED
,” “
BOLUSES FED
,” “
mL FLUSH
”) without clearing the pump settings, do the
following:
i. while
pump
RUNNING
ii. Press
HOLD
iii. Press
CLEAR VOLUME
iv. the pump will then show
LOAD A SET
or, if the Pump Set is already loaded, it will show
SET
LOADED
9.
KTO
: Keep tube open, this feature can be programmed to keep tube from occluding. Settings 5 – 240 minutes in
5 minute intervals. Approximately 1 mL fluid in 1 hour.
10.
Lock Feature
: While pump is running hold 3
rd
button until the lock icon appears
o
Unlock:
Hold 3
rd
button until lock icon disappears
11.
More:
Options
o
Buzzer/Alarm Buzzer increase or decrease volume
o
History/maximum of 72 hour past history
o
Languages/do not adjust
o
Continuous/Intermittent.
Most feeds in hospital will be Continuous
12.
Rinse out Bag
, use pump to rinse the bag. Use the
AUTO PRIME
or
HOLD TO PRIME
options.
Error/Warning/Information Screens and Troubleshooting
o
HOLD ERROR:
appears if the pump has been in
HOLD
mode, without input, for more than 10 minutes.
o
ROTOR ERROR
: appears during
RUNNING
or
PRIMING
, when the pump has detected an unusual
operating condition for the rotor. Generally results from a problem with the Pump Set tubing around the
rotor. Check that the Pump Set is not damaged, and re-load the tubing. Press
CONTINUE.
o
FLOW ERROR:
Downstream occlusion: occlusion in pump-patient line below pump
o
FEED OR FLUSH ERROR:
Upstream occlusion: Empty Bag or occlusion in bag-pump line above pump
o
PUMP SET DISLODGED:
Pump Set not properly loaded in pump
o
BATTERY LOW:
Indicates that the battery must be recharged immediately.
o
SYSTEM ERROR:
General Error caused by many factors: return pump to Biomedical Engineering
o
FEEDING COMPLETE
information screen appears after completion of the programmed feeding. Turn
pump off.
o
The Pump Set usage warning indicator (
SET USE >24 HRS
) will blink on the
RUNNING
screen if a Pump
Set has been used for 24 or more hours. It is recommended to replace Pump Sets after this length of
usage. This icon is only an informational message and does not require immediate action.
o
If troubleshooting reveals a problem with pump supplies (i.e. bag, tubing), complete product complaint
form and retain affected supplies to ensure follow-up by manufacturer