KANGAROO E-PUMP INSTRUCTIONS FOR FEED/FLUSH ONLY 

(adapted from Kangaroo ePump Instructions for Use Manual) 

 
 
 

CC.12.10A 

 BC Children’s Hospital Child & Youth Health Policy and Procedure Manual  

Page 2 of 2 

May 1

st

, 2012

                                     Refer to online version – Print copy may not be current – Discard after use

 

 

o

 Press 

RUN

 to restart pump at anytime 

 

8.  To clear feed totals (“

mL FED

,” “

BOLUSES FED

,” “

mL FLUSH

”) without clearing the pump settings, do the 

following: 

i. while 

pump 

RUNNING 

ii. Press 

HOLD 

iii. Press 

CLEAR VOLUME

 

 

iv.  the pump will then show 

LOAD A SET 

or, if the Pump Set is already loaded, it will show 

SET 

LOADED 

9. 

KTO

: Keep tube open, this feature can be programmed to keep tube from occluding. Settings 5 – 240 minutes in 

5 minute intervals. Approximately 1 mL fluid in 1 hour. 

 

10. 

Lock Feature

: While pump is running hold 3

rd

 button until the lock icon appears 

o

 

Unlock:  

Hold 3

rd

 button until lock icon disappears 

 

11. 

More:

 Options  

o

  Buzzer/Alarm Buzzer increase or decrease volume 

o

  History/maximum of 72 hour past history 

o

  Languages/do not adjust 

o

 Continuous/Intermittent. 

Most feeds in hospital will be Continuous  

12. 

Rinse out Bag

, use pump to rinse the bag. Use the

 AUTO PRIME 

or

 HOLD TO PRIME 

options.

 

 

Error/Warning/Information Screens and Troubleshooting 

o

 

HOLD ERROR: 

appears if the pump has been in 

HOLD 

mode, without input, for more than 10 minutes. 

o

 

ROTOR ERROR

: appears during 

RUNNING 

or 

PRIMING

, when the pump has detected an unusual 

operating condition for the rotor. Generally results from a problem with the Pump Set tubing around the 
rotor. Check that the Pump Set is not damaged, and re-load the tubing.  Press 

CONTINUE.

 

o

 

FLOW ERROR: 

Downstream occlusion: occlusion in pump-patient line below pump 

o

 

FEED OR FLUSH ERROR: 

Upstream occlusion: Empty Bag or occlusion in bag-pump line above pump 

o

 

PUMP SET DISLODGED: 

Pump Set not properly loaded in pump 

o

 

BATTERY LOW: 

Indicates that the battery must be recharged immediately.

 

o

 

SYSTEM ERROR: 

General Error caused by many factors: return pump to Biomedical Engineering 

o

 

FEEDING COMPLETE 

information screen appears after completion of the programmed feeding. Turn 

pump off. 

o

  The Pump Set usage warning indicator (

SET USE >24 HRS

) will blink on the 

RUNNING 

screen if a Pump 

Set has been used for 24 or more hours. It is recommended to replace Pump Sets after this length of 
usage. This icon is only an informational message and does not require immediate action. 

o

  If troubleshooting reveals a problem with pump supplies (i.e. bag, tubing), complete product complaint 

form and retain affected supplies to ensure follow-up by manufacturer 

 