Kangaroo® 224 Feeding Pump 

 

 

    Listed below are some of the probable cause of alarm conditions and their corrections. 

Symptom Probable 

Cause 

Corrections 

Flow Error 
Alarm 
(FLO Err) 

Feeding container is empty. 
 
Feeding tube or feeding set 
tubing is occluded. 
 
 
 
 
Drip chamber is improperly 
placed in pump. 
Drip chamber walls are coated 
with feeding formula. 
 
 
Sensors in upper drip chamber 
guide are blocked. 

Refill feeding container.  If bag has been in use for 
24 hours, replace. 
Locate point of occlusion and correct.  Note: 
occlusion may occur upstream or downstream from 
pump.   
Check to make sure roller clamp on feeding set 
tubing is open.  
Check to patency of the feeding tube. 
Check to make sure drip chamber is properly 
“locked” in position in drip chamber guide. 
Check to make sure formula is not preventing 
detectors from proper operation.  If formula cannot 
be removed from inside walls of chamber via 
chamber manipulation, replace set. 
Check to make sure detectors are free of dried 
formula.  Remove any deposits by using a cotton 
swab dampened with warm soapy water. 

Low Battery 
Alarm (Lo 
BAT) 

Battery has been run down 
below point of maintaining 
accuracy of pump. 

Plug pump power cord into wall outlet. 

Hold Error 
Alarm (Hld 
Err) 

Pump set has been left in the 
hold mode for over 2 ½ minutes. 

Press “START/HOLD” to silence alarm, then press 
“START/HOLD” again to restart pump. 

No Set (no 
SEt) 

Pump set has been improperly 
placed onto pump. 

Check to make sure only Kangaroo Pump Set has 
been placed onto the pump. 
Check to make sure that retainer clip on pump set has 
been properly positioned into retainer lock on pump. 

System  Error 
Alarm (SYS 
Err) 

Rotor is not turning while pump 
is running. 
 
Rotor is turning but is not 
detected by rotor sensors. 

Return pump for technical service. 
 
 
Return pump for technical service. 

No Audible 
Alarm 

Battery is fully discharged. 
 
Alarm circuitry failure. 

Charge battery for a minimum of 15 hours. 
 
Return pump for technical service. 

Troubleshooting Guide 