background image

Before use

3

Check

Particle filter is not dirty;  Water trap & flue probe empty of water;  

Water trap & red plug fitted to analyser;  Flue probe hose connected to flue gas inlet;  

Flue probe temperature plug connected to T1 temperature connection

 

Fresh air purge

Position flue probe in fresh air & press “On/Off”  

Analyser auto-calibrates within 60 seconds

Select “Ratio” on dial - in fresh air CO reading = 0ppm

Select “O

2

/Eff” on dial - in fresh air O

2

 reading = 20.9%

Select “Status” on dial to view

 

For best pressure sensor accuracy

Switch analyser on for 5 minutes to let temperature stabilise

Zero pressure sensor when analyser in exact position of use

 

To charge NiMH rechargeable batteries

Switch off

Connect charger - charging indicator illuminates

Display shows “BATTERY CHARGING” if batteries need charging

Connections

2

Battery charger

Watertrap drain 

with red plug

Flue probe 

temperature T1

Inlet temperature T2

Pressure 

connection P1

Pressure 

connection P2

- differential

Flue gas inlet

Red connector 

- plugs into flue 

gas inlet

Flue probe temperature plug

- plugs into T1

Narrower (+ve) pin must be on 

right hand side

Flue probe hose

Sensor upgrades 

available

Options

Sensor

Отзывы: