Page 32
APPENDIX 1 - MAIN PARAMETERS:
Here are the legends used and what they mean:
O
2
:
Oxygen reading in percentage (%)
O2R:
Oxygen reference setting. '----' means switched off or set to
0%.
T Flue:
Temperature measured by the flue gas probe in Centigrade. It
displays ‘
-
OC
-’
if the flue probe is disconnected.
T Inlet:
The temperature measure from the flue probe during the fresh
air purge will be used as the inlet temperature. If no probe is
connected then the internal ambient temperature of the
KANE251 will be used as the inlet.
T Nett :
Nett temperature calculated by deducting the
AMBIENT
or
INLET
temperature from the measured
FLUE
temperature.
Displays in either Fahrenheit (
F) or Centigrade (
C) or and
will display ‘
- - -
’ if the flue probe is not connected.
COp:
Carbon Monoxide
reading displayed in ppm (parts per
million). ‘
- - -’
is displayed if there is a fault with the CO
sensor or the instrument has not set to zero correctly, switch
off instrument and try again.
COm:
Carbon Monoxide displayed in mg/m
3
.
COn:
Carbon Monoxide in ppm normalised to the O2R value viewed
on the AUX screen
COM:
Carbon Monoxide in mg/m3 normalised to the O2R value
viewed on the AUX screen
R:
CO/CO
2
Ratio: measured CO divided by measured CO
2
CO
2
:
Carbon Dioxide calculation determined by fuel type. This is
only displayed when a combustion test is being carried out. ‘
-
- -
’ is displayed while in fresh air.