background image

12 KANE 458s MANUAL

USING YOUR KANE458s FOR THE FIRST TIME

Charge your analyser’s batteries for 8 hours – an overnight charge 

should be sufficient for an average 8-hour day.

Take time to read this manual fully & be aware your analyser’s  

configuration may not support all features detailed in this manual.

Take time to set up your requirements before using your analyser.

USING YOUR KANE458s EVERY TIME:

Check your analyser’s water trap is empty & particle filter is not dirty.

To empty, pull out the red plug, shake out water & replace red plug.

To change the particle filter, remove protective rubber cover, slide the 

water trap from analyser, remove particle filter from its spigot &  

replace. 

Reconnect your analyser’s water trap & rubber protective cover.

Connect your flue probe hose to your analyser’s flue gas inlet – see p.8.

Connect your flue probes temperature plug to your analyser’s T1 socket 

– check plug orientation is correct – See p.8.

 

FRESH AIR PURGE ON SWITCH ON

Your analyser starts a zero countdown to auto calibrate itself.

When finished, rotate dial to “RATIO” – In fresh air CO reading should be 

zero.

Rotate dial to “O2/EFF” – In fresh air O2 readings should be 20.9% + 

0.3%.

Содержание 458s

Страница 1: ...Stock No MAN00099 Rev 0 00000 JULY 2020 Kane International Ltd KANE 458s Flue Gas Analyser with direct CO measurement and CO sensor protection...

Страница 2: ...FOR THE FIRST TIME 12 13 USING YOUR KANE458S EVERY TIME 12 FRESH AIR OURGE ON SWITCH ON 12 13 USING THE MENU 14 16 MEASURING FLUE GASSES 16 CO PROTECTION PUMP OPERATION 17 USING YOUR ANALYSER 18 19 C...

Страница 3: ...ROOM CO DISPLAY 28 USING YOUR KANE458s AS A THERMOMETER OR PRESSURE METER 27 28 ANALYSER PROBLEM SOLVING 29 SPECIFICATIONS 30 31 EU DECLARATION OF CONFORMITY 32 SERVICE CALIBRATE RECERTIFY 33 37 KANE...

Страница 4: ...ir during initial countdown to protect you from unsafe levels of CO and CO2 possibly present in the room Depending on your options these parameters are calculated Oxygen O2 CO CO ratio Carbon Monoxide...

Страница 5: ...our optional KANE IRP 2 infrared printer or KANE s wireless printer App The memory can store up to 60 combustion tests 20 AUX tests 20 let by tightness tests 20 temperature pressure tests 20 room CO t...

Страница 6: ...6 KANE 458s MANUAL ANALYSER FEATURES AND KEYPAD...

Страница 7: ...int report short press to enter a print or wireless data transfer Navigate up short press to scroll up Enter key used to select current option Navigate down short press to scroll down Data hold short...

Страница 8: ...re T1 Inlet temperature T2 Pressure connection P1 Pressure connection P2 differential Flue gas inlet On Off Function keys see below Water trap particle filter Rotary switch Red connector plugs into Fl...

Страница 9: ...9 KANE 458s MANUAL Watertrap drain with red plug Thermocouple Stainless Steel Shaft Depth Stop Cone Gas Path Magnets x2 Thermocouple wire...

Страница 10: ...is correct Replace battery cover protective rubber cover TIME AND DATE After changing the batteries reset your analyser s time date CHARGING NIMH BATTERIES Your KANE458s uses a standard Micro USB conn...

Страница 11: ...so exposed to a known gas mixture but is allowed to settle to a steady figure and the reading adjusted to the stated gas concentration of the test gas Protection Against Electric Shock In accordance w...

Страница 12: ...pull out the red plug shake out water replace red plug To change the particle filter remove protective rubber cover slide the water trap from analyser remove particle filter from its spigot replace Re...

Страница 13: ...WARNING This message says your analyser must be reset in fresh air Place your analyser in outdoor fresh air press You can perform a manual ZERO any time rotate dial to RATIO hold down the button to se...

Страница 14: ...rotate dial to any position any changes not entered will not store As you scroll up or down the side LEDs illuminate to point to the active line MAIN MENU SUB MENU OPTIONS COMMENTS SETUP LANGUAGE ENG...

Страница 15: ...f stabilisation in minutes Default 1 minute PASS KEY TIGHTN S Set duration of tightness test in minutes Default 2 minute BACK MAIN MENU SUB MENU OPTIONS COMMENTS UNITS FUEL TYPE NAT GAS TOWN GAS COKE...

Страница 16: ...nto the appliance s sampling point The tip of the probe should be at the centre of the flue Use the flue probe s depth stop cone to set the position With balanced flues make sure the probe is position...

Страница 17: ...tions are achieved readings are within expected range CO PROTECTION PUMP OPERATION Your analyser s CO sensor is automatically protected from high levels of CO When levels of CO rise above 2000ppm the...

Страница 18: ...dial to Ratio to display RATIO DISPLAY Fuel type can be changed via Menu or STATUS setting CO CO2 ratio Carbon monoxide ppm Carbon dioxide Excess air Pressure reading Press button to send a full comb...

Страница 19: ...ature oC normally set by flue probe during fresh air purge NO sensor not fitted Differential temperature NO sensor not fitted TRANSFERING RESULTS To print simply press and release the button to send r...

Страница 20: ...stop flashing To view logged data press or to highlight another line Press sidelights will flash on that line Use or to scroll through data To finish press Sidelights stop flashing Use or to scroll d...

Страница 21: ...21 KANE 458s MANUAL Auxiliary Commission Combustion Let by Tightness Type A Pressure Temp Sweep Test PRINTOUTS...

Страница 22: ...l to 0 20 TEST 2 Insert your flue probe into the appliance s exhaust outlet to measure CO CO RATIO levels these must be within the manufacturer s instructions If manufacturer s instructions are not av...

Страница 23: ...resolution readings can be selected via Menu Pressure units can be selected via Menu Eg Flow Temp Eg Return Temp Differential Temp PRESSURE TEMPERATURE TESTING WARNING Never attempt to take a pressur...

Страница 24: ...ns thoroughly If in doubt contact the manufacturer After adjusting a gas air ratio valve you must ensure CO CO CO CO ratio readings are within the manufacturer s specified limits LARGE BORE TUBING ISS...

Страница 25: ...to start the let by test display shows If let by test fails rotate dial to another position to stop the test If Let by test passes adjust gas pressure for the tightness test press to start the stabil...

Страница 26: ...ains how to view print stored reports Tightness test is automatically stored in the memory Pressure at start of tightness test Real time pressure reading Tightness default time is 120 seconds can be c...

Страница 27: ...lts stored every minute LIMIT 10ppm ALARM 30 ppm TYPE B BOILER OPEN FLUE 15 minute test with results stored every minute LIMIT 10ppm ALARM 30 ppm TYPE A COOKER 30 minute test with results stored every...

Страница 28: ...m CO test at any time by pressing Otherwise Room CO tests automatically end after the pre set time Room CO tests are automatically stored in your analyser s memory as a log number You can send your Ro...

Страница 29: ...Particle filter blocked Probe or tubing blocked Pump not working or damaged with contaminants Net temperature or Efficiency calculation incorrect Ambient temperature set wrong during Automatic Calibr...

Страница 30: ...Battery Life 8 hours continuous with pump on Certification The KANE458s is independently tested and certified to EN 50379 Parts 1 3 in accordance to 1st German Federal Emission Control Ordinance Bim5c...

Страница 31: ...GE RESOLUTION ACCURACY Operating Conditions Temperatures 0 45 C Humidity 15 to 90 RH non condensing Power Supply Rechargeable batteries USB Charging Physical Characteristics Weight Approx 0 625g Dimen...

Страница 32: ...01165EU Restriction of the use of certain hazardous substances in electrical and electronic equipment RoHS The following harmonised standards and technical specifications have been applied CERTIFICATI...

Страница 33: ...BRATE RECERTIFY All analysers pressure meters should be recertified annually Extend your KANE analyser and pressure meter s no quibble warranty up to 10 years by returning your analyser pressure meter...

Страница 34: ...try colleagues unknowingly buying stolen goods Remove your KANE Analyser once sold so its new owner can also benefit Excludes KANE 9 series analysers UKAS certificates The fastest way to manage your a...

Страница 35: ...35 KANE 458s MANUAL KAM Your support our way...

Страница 36: ...o uk Select FREEPOST for tracked carriage UK mainland only Your analyser will be despatched on the same day we receive it Book pay to Service Recertify via your KAM dashboard OR YOUR MONEY BACK GUARAN...

Отзывы: