background image

 

KANE455 manual 

Page 46

 

Why did you buy a Kane Product? 

 

…

  

Dealer Recommendation

 

…

  

Other Recommendation

 

…

  

Value for Money

 

 

…

  

Our Fixed Price Servicing Programme

 

…

  

Kane

 

 

 

 

…

  

Previous Owner

   

…

  

Not your Decision

 

 

…

  

Other:

 

 
 

What brand was your previous analyser? 

 
 

How did you hear about Kane? 

 

…

  

Magazine Advert

 

 

 

…

  

Trade Counter Literature

 

 

…

  

Personal Recommendation

   

…

  

Internet

 

…

  

Exhibition

 

 

 

 

…

  

Other:

 

 
 

Which do you read most often? 

 

 

 

 

 

Often

  

Sometimes

   

Hardly Ever

 

Registered Gas Engineer

 

   

…

 

 

       

…

   

 

      

…

 

Gas Installer

 

 

 

   

…

 

 

       

…

   

 

      

…

 

P.H.P.I.

 

 

 

 

   

…

 

 

       

…

   

 

      

…

 

P.H.A.M. News

 

 

 

   

…

 

 

       

…

   

 

      

…

 

Heating Ventilating & Plumbing

    

…

 

 

       

…

   

 

      

…

 

Heating & Plumbing Monthly

 

   

…

 

 

       

…

   

 

      

…

 

 
 
 
 
 
 

Thank you for completing this survey. 

All the information we have collected is confidential. 

We do not sell or share data with any other company or organisation. 

 

For latest prices and delivery to your door visit MyTub Ltd - www.mytub.co.uk - [email protected]

For latest prices and delivery to your door visit MyTub Ltd - www.mytub.co.uk - [email protected]

Содержание 455

Страница 1: ...urement Stock No 18859 7 October 2012 Kane International Ltd For latest prices and delivery to your door visit MyTub Ltd www mytub co uk info mytub co uk For latest prices and delivery to your door vi...

Страница 2: ...TIAL TEMPERATURE 19 4 5 ROOM CO TESTING 20 4 6 KANE455 PRINTOUTS 21 5 USING THE MENU 22 23 6 USING THE KANE455 AS A THERMOMETER OR 24 25 PRESSURE METER 7 MEASURING FLUE GASES 26 8 ANALYSER PROBLEM SOL...

Страница 3: ...TERS 38 40 ADDENDUM OPTIONAL NITRIC OXIDE NO SENSOR 41 44 For latest prices and delivery to your door visit MyTub Ltd www mytub co uk info mytub co uk For latest prices and delivery to your door visit...

Страница 4: ...ands free operation and is supplied with a flue probe with integral temperature sensor The large display shows 4 readings at a time and all data can be printed via an optional infrared printer The pri...

Страница 5: ...nter Menu controls Scroll up down Tasklight and infra red emitter Flue Gas Inlet Temp and pressure connections Battery Compartment behind rubber sleeve For latest prices and delivery to your door visi...

Страница 6: ...455 manual Page 6 For latest prices and delivery to your door visit MyTub Ltd www mytub co uk info mytub co uk For latest prices and delivery to your door visit MyTub Ltd www mytub co uk info mytub co...

Страница 7: ...H Batteries Ensure that you use the correct charger The part number is KMCU250 UK To fully charge NiMH batteries Switch the KANE455 on The charger must then be connected and switched on When charging...

Страница 8: ...rrect see Page 6 2 1 FRESH AIR PURGE Position the flue probe in fresh air then press The analyser s pump starts and the analyser auto calibrates for approximately 90 seconds When complete Select Ratio...

Страница 9: ...e used in well ventilated locations by trained and competent persons after due consideration of all the potential hazards Users of portable gas detectors are recommended to conduct a bump check before...

Страница 10: ...ature If an inlet temperature probe is not connected to the analyser during countdown the measured temperature from the flue probe will be used as the inlet temperature If neither probe is connected d...

Страница 11: ...will automatically switch itself off when the rotary switch is set to Menu Status Pressure Tightness or Differential Temperature Zeroing the pressure sensor To re zero the pressure sensor when Prs is...

Страница 12: ...eration if the pump is switched off Using Buttons The function buttons below the symbols are used to navigate through the menu when the rotary switch is set to MENU See USING THE MENU Section 5 For la...

Страница 13: ...3 Carbon dioxide Press to print a full combustion test or send to PC via optional Bluetooth Hold for 2 seconds to log a full combustion report O2 EFF display O2 9 8 Oxygen left after combustion Should...

Страница 14: ...ull combustion test or send to PC via optional Bluetooth Hold for 2 seconds to log a full combustion report Viewing printing a logged combustion test Select MENU REPORT COMB N VIEW Use and to select t...

Страница 15: ...l Bluetooth Hold for 2 seconds to log a pressure report Viewing printing a logged pressure test Select MENU REPORT PRESSURE VIEW Use or to select the log number to be printed Press to print the test o...

Страница 16: ...range tube over the rim of the spigot to ensure a gas tight seal This may not produce a gas tight seal For latest prices and delivery to your door visit MyTub Ltd www mytub co uk info mytub co uk For...

Страница 17: ...P2 10 15 Real time pressure reading TIME 59 Let by default time is 1 minute Can be changed via Menu If the let by test fails simply move the rotary switch to any position other than tightness to abor...

Страница 18: ...ss test Select MENU REPORT TIGHTN S VIEW Use or to select the log number to be printed Press to print the test or send to PC via optional Bluetooth Note The analysers s memory can store up to 20 tight...

Страница 19: ...time temperature difference Press to print a differential temperature test or send to PC via optional Bluetooth Hold for 2 seconds to log a differential temperature report Viewing printing a different...

Страница 20: ...can stop the Room CO test at any time by pressing If not stopped earlier the Room CO test will automatically end after 30 minutes The CO test series is automatically stored in the memory as a log numb...

Страница 21: ...IR 34 8 CO CO2 0 0001 PRS mbar 0 00 K455 1 0 YOUR COMPANY NAME PHONE NUMBER HERE K455 1 0 YOUR COMPANY NAME PHONE NUMBER HERE ROOM CO TEST Customer LOG 04 TIME 11 53 15 05 07 LOG 01 Appliance TIME 12...

Страница 22: ...MM YY format EXIT PRESSURE SMOOTH OFF normal response ON slower damped response RESOLVE LOW e g 0 01mBar resolution HIGH displays to an extra decimal place PS UNITS mBar mmH2O Pa kPa PSI mmHg hPa InH...

Страница 23: ...on the AUX display LINE 1 LINE 2 LINE 3 LINE 4 EXIT HEADER Printout header 2 lines 20 characters per line HEADER 1 HEADER 2 EXIT EXIT SERVICE CODE Password protected for authorised service agents onl...

Страница 24: ...Use the T1 connection for the flow temperature sensor T2 21 3 Use the T2 connection for the return temperature sensor T 0 0 Real time temperature difference The standard printout for this mode is as f...

Страница 25: ...range tube over the rim of the spigot to ensure a gas tight seal This may not produce a gas tight seal For latest prices and delivery to your door visit MyTub Ltd www mytub co uk info mytub co uk For...

Страница 26: ...er s internal temperature operating range Do not put the analyser on a hot surface Do not exceed the water trap s levels Do not let the analyser s particle filter become dirty and blocked View the dis...

Страница 27: ...tput Fuse blown in charger plug Analyser does not respond to flue gas Particle filter blocked Probe or tubing blocked Pump not working or damaged with contaminants Net temperature or Efficiency calcul...

Страница 28: ...lly charged NiMH batteries fitted This is not a problem and is to be expected NiMH batteries only deliver 1 25 V per cell whereas Alkalines deliver 1 5 V per cell Fresh alkalines give a BAT value of 9...

Страница 29: ...ton near Manchester Tel 01942 873434 the primary service centre for UK KANE455 customers and at Welwyn Garden City in Hertfordshire Tel 01707 375550 the primary service centre for non UK customers By...

Страница 30: ...250 0 0 999 0 1 1ppm 0 1 0 1 0 1 0 1 0 0001 0 3 1 3ppm 5 reading 0 3 reading 1 0 reading 1 0 reading 0 2 reading 5 reading Pressure differential Nominal range 80mBar Maximum over range without damage...

Страница 31: ...rger for NiMH batteries only Dimensions Weight Handset Probe 0 8kg handset with protective cover 200 x 45 x 90mm 300mm long including handle 6mm diameter x 240mm long stainless steel shaft with 3m lon...

Страница 32: ...mize interference or switch off if possible the offending equipment during your test At the time of writing this manual December 2009 Kane International Ltd are not aware of any field based situation...

Страница 33: ...60 Hz with a current capability of 0 3 A The chargers output voltage is 9 V dc at a maximum of 0 66A The charger has no user serviceable components Only a correctly specified and rated charger must be...

Страница 34: ...recommended that the analyser is given the protection of a carry case during transportation it is not required for normal operation k The KANE455 has an initial start up delay following switch on of a...

Страница 35: ...e emptying of the water trap are given in Section 2 of this manual m WARNING When using a KANE455 to test an appliance a full visual inspection of the appliance in accordance with its manufacturer s i...

Страница 36: ...455 manual Page 36 For latest prices and delivery to your door visit MyTub Ltd www mytub co uk info mytub co uk For latest prices and delivery to your door visit MyTub Ltd www mytub co uk info mytub c...

Страница 37: ...455 manual Page 37 For latest prices and delivery to your door visit MyTub Ltd www mytub co uk info mytub co uk For latest prices and delivery to your door visit MyTub Ltd www mytub co uk info mytub c...

Страница 38: ...ternal ambient temperature will be used as the inlet temperature T Nett Nett temperature calculated by deducting the INLET temperature from the measured FLUE temperature It displays OC if the flue pro...

Страница 39: ...inted or stored Note When changing the batteries on the instrument the memory will store the date and time for up to one minute if outside this time it may be necessary to re enter the details Date an...

Страница 40: ...y EfN Nett efficiency EfC Condensing efficiency PO Pump off Calculated oxygen greater than 18 so calculation is disabled OC Open circuit temperature input CAL Number of days left before recalibration...

Страница 41: ...KANE455 analysers fitted with optional Nitric Oxide NO sensors For latest prices and delivery to your door visit MyTub Ltd www mytub co uk info mytub co uk For latest prices and delivery to your door...

Страница 42: ...customise the parameters on the AUX display LINE 1 LINE 2 LINE 3 LINE 4 EXIT HEADER EXIT SERVICE BLUE COM Bluetooth is a factory fitted optional extra NOTE To EXIT the MENU at any time simply move the...

Страница 43: ...our lines and then EXIT Rotate the dial from MENU to AUX to display all your chosen settings PRINTING and STORING The NO reading are printed and stored in the same way as the other combustion gas read...

Страница 44: ...solution 1 ppm 1ppm Accuracy 2ppm 30ppm 1 5 ppm 30ppm 5ppm 100ppm 1 5 reading 100ppm Range 0 to 100 ppm 0 to 1000 ppm 1 Using dry gases at STP For latest prices and delivery to your door visit MyTub L...

Страница 45: ...2 Town City County Postcode Country Phone Number Fax Number Mobile Number Email Address Product Details Note Proof of Purchase may be required for warranty claims Date Purchased as numbers 05 01 10 P...

Страница 46: ...Exhibition Other Which do you read most often Often Sometimes Hardly Ever Registered Gas Engineer Gas Installer P H P I P H A M News Heating Ventilating Plumbing Heating Plumbing Monthly Thank you fo...

Страница 47: ...455 manual Page 47 For latest prices and delivery to your door visit MyTub Ltd www mytub co uk info mytub co uk For latest prices and delivery to your door visit MyTub Ltd www mytub co uk info mytub c...

Страница 48: ...ster on our website www kane co uk or complete detach and return the Product Registration page For latest prices and delivery to your door visit MyTub Ltd www mytub co uk info mytub co uk For latest p...

Отзывы: