BEFORE USING ANY ANALYSER CHECK THAT: 

 

-  The calibration date has not expired 
-  The particle filter is not dirty 
-  The water trap and flue probe hose are empty of water 
-  The water trap is fitted correctly to the instrument including the drain plug 
-  The flue probe hose is connected properly to the water trap  
-  The flue temperature plug is connected into Tf 

 

USING THE ANALYSER 

 

Switch the analyser on.  It will now purge for approximately 20 seconds after which 
it will show a set a readings on the display and now is a good time to check the 
battery status. 
 
Using the               key, scroll until the display shows the CO/CO

2

 ratio or whatever 

you want to look at. 
 
Insert the probe into the centre of the flue after approximately 60 seconds the 
readings should stabilise. 
 

STORING A TEST 

 

Press the               key once the display will show you at what number it will record 
your test. 
 
Press the               key again and it will store your test. 
 

RETRIEVING A TEST 

 

Press the              key once, then press the              key.  View will now be shown 
on the screen. 
 
Using the               key you can now count up to the number where your test was 
stored.   

 

When you’ve reached the correct number, press the             key and your test 
readings will be displayed. 

 

PRINTING A TEST 

 

You can now  print the test by pressing the              key. 
 
To return to the main screen, press the          key. 

 

Remove the probe from the flue and allow the analyser to purge with fresh air 
before switching off! 