KSP
‐
F01A
Dosing
Pump
User
Manual
11
5
Flow
rate
adjustment
Only
in
Manual
mode,
you
can
adjust
dosing
flow
rate
through
turning
the
backside
rotary
knob
of
the
dosing
pump.
Note:
You
can
change
the
flow
rate
only
in
manual
mode.
In
other
states,
the
rotary
knob
does
not
change
the
flow
rate
immediately.
The
next
time
you
start
the
manual
run,
the
dosing
pump
will
run
according
to
the
adjusted
position.
6
Calibration
6.1
Overview
Calibration
is
a
process
to
determine
accurate
flow
rate
by
using
a
proper
measuring
tool
to
measure
added
volume
during
a
certain
time
period.
(10ml
measuring
cylinder
is
delivered
for
free
with
the
machine,
for
more
accurate
flow
rate,
a
larger
measuring
cylinder
can
be
used).
Pump
is
needed
to
be
calibrated
under
the
following
occasions:
Dosing
pump
used
for
the
first
time;
After
running
for
a
long
time,
the
displayed
flow
rate
is
quite
different
from
actual
flow
rate
(pump
may
be
wore);
Flow
rate
is
adjusted
by
using
rotary
knob.
6.2
Operation
1.
connected
to
the
pipeline;
2.
Empty
the
liquid
in
the
pipeline;
3.
Empty
the
graduated
cylinder;
4.
Press
the
"calibration"
key
to
start
calibration;
5.
Waiting
for
the
end
of
the
countdown
operation;
6.
Measure
the
volume
of
liquid
in
the
cylinder
7.
Input
the
volume
by
press
and