User Manual
OBM
14
When cycle has started you see this screen for automatic program :
Bath temperature:
current measured bath temperature is
26.31
°C
Heating power:
current heater duty cycle
0%
(from
0%
to
100%
)
Temperature status:
Stabilizing
= temperature is stabilizing and it is not yet in setpoint tolerances
Stable
= temperature is stable and it is in setpoint tolerances
Current program:
the current running program number is
2
(
0
- manual program,
1
÷
10
for automatic
program)
Setpoint temperature:
current setpoint temperature
50.00
°C
(setpoint range from
35
.00°C
to
220
.00°C
)
Stirrer speed:
the set speed for stirrer is
100%
( set range from
30%
to
100%
)
Step dwell time:
remaining time of temperature stability for current step
is
0:30:00
= 30 minutes
(in hh:mm:ss format). Set range from 00:01 to 24:00 (in hh:mm format).
For endless dwell time set program to 00:00.
Step information:
1/ 1
( current temperature step / all temperature steps in program).
Maximum number of temperature steps is 10.
When cycle has finished you see confirmation screen:
Setpoint temperature
Heating power
Stirrer speed
Bath temperature
Current program
Temperature status
Step information
Step dwell time