6
insulation may result, which could cause current leakage or electric
shock.
CAUTION!
To allow for proper air flow, a minimum of fifteen centimeters clearance is
required behind the freezer.
5.2 Wiring
CAUTION!
Connect the equipment to the correct power source. Incorrect voltage can
result in damage.
CAUTION!
For personal safety and trouble-free operation, the unit must be properly
grounded. Failure to ground the equipment may cause personal injury or
damage to the equipment. If the power supply outlet is not grounded, it is
necessary to install a ground from the grounding screw located on the lower
right rear side of the unit.
If possible, it is preferable to connect the freezer to a dedicated (separate) power
supply. Each freezer is equipped with a service cord and plug designed to connect it to
a power outlet which it delivers the correct voltage. Supply voltage must within
+10% to -10%
of the freezer
’
s rated voltage.
CAUTION!
Never cut, or remove the grounding prong from the service cord plug. If the
prong is removed, the freezer will not be properly grounded.
5.3 Leveling
The unit must be level: Both front to back and side to side.
To level chest models, leave the casters in place and shim the wheel(s) with strips of
sheet metal cut wider than the caster.
Upright models have adjustable leveling feet located adjacent to the front casters.
Extend the leveling feet by rotating them counterclockwise. Ensure the unit is100%
level.