Please read the following precautions carefully to ensure the safety of the patient and the operator 

when using the probes.

 

Warning:

 

13.

 

Adapter has no switch. APPLIANCE COUPLER or MAINS PLUG is used as the 
isolation means from the SUPPLY MAINS. Not to position the equipment so that 
it is difficult to operate the disconnection device. 

14.

 

Other  additional  multiple  socket-outlets  or  extension  cords  should  not  be 
connected to the equipment. 

15.

 

When the equipment works abnormally, do stop working, turn off the power and 
check the reason, then contacts the Kaixin Company about it. 

16.

 

It is forbidden to load and unload the probe or move the equipment in galvanic to 
avoid danger of safety. 

17.

 

The  equipment  is  switched  completely  only  by  disconnecting  the  power  supply 
from the wall socket. 

18.

 

Turn  off  power  and  pull  out  of  the  plug  from  socket  after  the  equipment  is 
completed operation. 

19.

 

It  is  forbidden  to  drag  and  press  the  power  and  probe  cables  emphatically; 
regularly  inspect  whether  there  is  pull-apart  and  bareness,  if  there  is  the 
phenomena like this, turn off power supply immediately, stop using it and change 
it for new one. 

20.

 

Pull out of the plug from socket after operation in thunderstorm weather to avoid 
the equipment being damaged by lightening. 

21.

 

Always keep the equipment dry. Avoid transporting this equipment quickly from 
a cold place to a warm place; otherwise condensation or water droplets may form 
allowing a short circuit and possible electric shock. 

22.

 

To avoid damaging power adapter or harming people by unexpected fallen, make 
sure the power adapter is placed on the leveled desk. 

23.

 

The  operator  must  not  touch  signal  input/  signal  output  and  patient 
simultaneously. 

Attention:

 

1.

 

Please keep the equipment in a frozen state when it is turned on and not in use to 
extend the service life of the probe.

 

2.

 

Repeat available reboot time should be more than one minute to avoid turn on/off 
power supply in short time.

 

3.

 

Using radio transmitting equipment nearby the equipment may interfere with the 
normal  operation  of  the  equipment.  Prohibited  carry  or  use  of  devices  that  can 
generate  radio  waves  within  the  room  installed  this  equipment,  such  as  cell 
phones, radio transceivers and wireless remote control toys.

 

4.

 

Do not apply  external  force to  the  equipment; otherwise, the equipment  may be 
damaged.

 

5.

 

If the equipment is used in a small room, the room temperature may rise. Please 
provide proper ventilation and free air exchange.

 

6.

 

The  images  displayed  in  this  equipment  are  only  reference  for  diagnosis.  Our 
company  is  not  responsible  for  the  correctness  of  diagnostic  results.  The  doctor 
shall be responsible for the diagnostic process.

 

Содержание KX5600F

Страница 1: ...KAIXIN KX5600F Full Digital Ultrasonic Diagnostic Instruments Vet User s Manual Xuzhou Kaixin Electronic Instrument Co Ltd...

Страница 2: ...Symbols Description 3 2 Chapter 4 Installation and Check 4 1 4 1 System Installation 4 1 4 2 Ultrasonic Probe Installation 4 1 4 3 Shutter Release Installation 4 2 4 4 Video Recorder Installation 4 3...

Страница 3: ...ce 7 1 7 1 Daily Maintenance 7 1 7 2 Use and Maintenance for the Rechargeable Lithium Ion Battery Pack 7 3 7 3 Replaceable Parts 7 5 7 4 Backup Data 7 5 7 5 Troubleshooting 7 5 7 6 Periodic Safety Che...

Страница 4: ...ng requirements 1 Assembly expansion readjustment improve and repair are all performed by professionals recognized by Kaixin 2 All replacement parts and accessories consumables involved repairs are Ka...

Страница 5: ...lteration or repair the system by non Kaixin designated personnel 6 This product is intended to provide clinical diagnostic data for the doctor The doctor shall be responsible for the diagnostic proce...

Страница 6: ...result in light injury or property loss or affecting the use if not avoided Safety Symbols Symbols Description Type B applied part Indicates the need for the user to consult the instructions for use f...

Страница 7: ...the equipment please immediately turn off the power and contact your service representative 6 Do not use a probe that has a damaged or scratched surface Immediately stop using the probe and contact yo...

Страница 8: ...pment being damaged by lightening 21 Always keep the equipment dry Avoid transporting this equipment quickly from a cold place to a warm place otherwise condensation or water droplets may form allowin...

Страница 9: ...o practice ALARA when operating ultrasound examination Minimize the acoustic power without compromising the quality of images 7 Clean and disinfect the probe before and after each examination 8 Thorou...

Страница 10: ...c pressure 700hPa 1060hPa 1 3 Dimensions and Weight 1 Main unit dimension approx 285 143 35mm3 L W T 2 Main unit net weight 1 1kg excluding accessories Warning Do not use or store the equipment in the...

Страница 11: ...1 PC 4 Power cable 1 PC 5 Shutter release 1 PC 6 Probe holder 1 PC 7 Strap 1 PC 8 Leather bag 1 PC 9 Plastic sealed box 1 PC 2 2 Optional Parts 1 3 5 MHz convex array probe 2 3 3 5MHz linear array lo...

Страница 12: ...o ultrasonic signal based on principle of converse piezoelectric effect The ultrasonic signal after entering the human body is reflected as echo wave and converted to electric signal again The acousti...

Страница 13: ...ates 9 Power switch 3 5 Symbols Description This instrument uses the symbols listed in the following table and their descriptions are explained Symbols Description Type B applied part Indicates the ne...

Страница 14: ...acking limit by number Temperature limits Storage and transport Humidity limitation Storage and transport Atmospheric pressure limitation Storage and transport Marking for the separate collection of e...

Страница 15: ...the ultrasonic system before disconnecting the probe Disconnecting the probe with system power on may damage the system or probe 2 Before disconnecting the probe place the cable and probe on a stable...

Страница 16: ...direction indicated on the label Before turning the ring clockwise be sure to push the probe connector into place and then rotate to prevent the ring from idling Probe connection schematic 4 2 2 Ultra...

Страница 17: ...face of the main unit 2 Connect to the main power supply Insert the power plug jack furnished with the machine into power input socket of the power adapter the other end to the mains socket outlet The...

Страница 18: ...f the machine for possible damage or blockage 10 Clean the field and environment 4 7 2 Inspection After Start up Check the following items after starting the machine 1 No abnormal voice strange smell...

Страница 19: ...Day Hour and Min 4 When setting year month day hour and minute press direction key to increase value or press direction key to decrease value 5 Press key to confirm the time setting and quit setting i...

Страница 20: ...D E F G H I J K L M Shift 5 6 7 8 9 N O P Q R S T U V W X Y Z Press direction keys to move cursor to point to Caps and then press key to achieve capital and small letter conversion If the cursor poin...

Страница 21: ...5 3 8 WIFI Switch Setting 1 Press key to enter main menu press direction keys to move cursor to Setup 2 Press key to enter setting interface 3 Press direction keys to move the symbol to point to WIFI...

Страница 22: ...s to move cursor to M Mode and then press direction keys to change the M scan mode 3 Change M scan speed Press key to enter main menu press direction keys to move cursor to M Speed and then press dire...

Страница 23: ...ion keys to move cursor to Dyn in the display area Press direction key to increase the value of dynamic range and direction key to decrease the value of dynamic range so as to control the dynamic rang...

Страница 24: ...ys to select L or R to achieve image left right reverse The image is flipped left right that is the probe scanning direction is changed The probe scanning direction is indicated by the arrow on the up...

Страница 25: ...time status Note As the swine increases weight and age and if the measurement site is accurate two or three layers of backfat thickness can be measured the backfat measurement values for two or three...

Страница 26: ...e image code need to be saved such as choose 003 4 Press key the current image is saved in the frame for coded 003 The saved image code is preceded with asterisk 5 Press key to quit saving status and...

Страница 27: ...key to enter manual playback status in the manual playback status press direction keys to view images frame by frame continuously press key three times again to return to automatic playback status In...

Страница 28: ...s and retype 5 13 Check List Management 5 13 1 Save and View Check List 1 Freeze the desired image 2 Record the ID and Result according to the method in section 5 11 Text Input press key to confirm an...

Страница 29: ...6 to complete the multi group data measurement 8 Continuously press key twice to quit the measurement status 9 Press direction keys to choose Clear in display area and press key to clear all marks and...

Страница 30: ...trace on the screen trace is automatically drawn along the order of marking point until close the beginning point and end point the measured circumference and area values are automatically displayed a...

Страница 31: ...of the measurement 5 Press direction keys to move the mark to the end point of the measurement at the same time a lighted dotted line appears between the starting point and the end point as the dashe...

Страница 32: ...s displayed in real time on the right part of the screen measure loin with distance measurement method the loin is displayed in real time on the right part of the screen press key to finish the loin m...

Страница 33: ...performed by professional trained engineer or contact the Kaixin Company 2 For the system performance and safety you should perform periodical checks for the system 3 The following maintenance should...

Страница 34: ...se or soak the probe connector or close to connector cable The soaking time of probe in the disinfectant is limited to the minimum time recommended by disinfectant manufacturer e g Cidex OPA manufactu...

Страница 35: ...ower adapter cable for no wrinkles cracks or aging and check the surface of the power adapter free from cracks and expansion 2 Manual operation check the power plug connection is reliable no loose or...

Страница 36: ...t capability or different model or different brand battery which will lead to the leakage explosion or fire 13 Do not use the abnormal battery with particular smell or abnormal heat or distortion or t...

Страница 37: ...and the screen does not display when starting the machine 1 Check power supply 2 Check power supply cord and plug 3 Check power adapter 2 Character and gray scale are displayed but no ultrasonic image...

Страница 38: ...ing 1 year When check the instrument please consult the International Trade Dept of Kaixin or its dealers as they need to have professional technology engineers Inspect sorts Inspect items Cleaning Sy...

Страница 39: ...instrument case should have adequate space from ground walls and ceiling of the warehouse 2 Environment requirement Ambient temperature 20 55 Relative humidity 30 93 without condensation Atmospheric...

Страница 40: ...eration in the presence of a flammable anesthetic mixture with air or oxygen or nitrous oxide Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or oxygen or nit...

Страница 41: ...n Mansion C 01 Economic Development Zone Xuzhou Jiangsu China Post Code 221004 Tel 86 516 87732932 87733758 Fax 86 516 87732932 87792848 Website http www kxele com E mail info kxele com Information co...

Отзывы: