Page 5 of 101
outlet with the isolation transformer, you should connect the plug of the system to the
hospital-graded standard socket. Meanwhile, please consult the professional technician to ensure that
the connection meets the safety requirements.
The video printer should be connected to the cable provided by the manufacturer, otherwise, there is a
danger of electric shock.
It is recommended to connect this equipment to equipotential system. Use yellow and
greenequipotential grounding cables, one end is connected to position with symbol, and other
end is connected to equipotential system. Use of potential equalization conductor together with
areference to requirements of IEC 60601-1 for Medical Electrical System.
Do not pour any fluid onto the ultrasound system surfaces, as fluid seepage into the electrical circuitry may
cause excessive leakage current or system failure. If carelessly pour any water onto the system,
immediately stop using the ultrasound system and contact Service Representative immediately.
Only use the probes provided by the manufacturer. Otherwise, the ultrasound system cannot be
performed, and an accident such as a fire may result in the worst case.
The machine that are not serviced or maintained while in use with the patient.
Make sure the system is powered off and power cable is disconnected before cleaning the system.
Otherwise, an electric shock may happen.
Do not use this system in the strong electromagnetic field. Using the system in the improperenvironment may
result in malfunction or damage.
Only the peripherals and accessories (such as probes, peripherals or cables) provided or recommended by
the manufacturer can be used. Using other devices or accessories may degrade the system performance
and even cause an electrical shock.
1.1.2 Mechanical Safety
Please make sure that the system is powered off and the power cable and the relevant accessories are
disconnected before moving or transporting the system.
Place the system on a level desk.
The plug is used as disconnect to the mains supply, do not to position the machine so that it is
difficult to operate the disconnection device.
Do not place the system on a tilted plane with the angle larger than 10
°
. Otherwise, the systemwill
fall off to cause system damage or personal injury.
1.1.3 Accessories Safety
Use the probe carefully. If any part of the transducer surface is scratched, immediately stop using the probe.
Otherwise, there is a danger of electric shock.
After disinfecting the accessories, chemicals must be washed out from the accessories. Remaining