Kailiu Instrument MGG/KL-DCB Скачать руководство пользователя страница 10

7

signal output. Short measurement filtering time shows fast corresponding speed.
Flow unit: Instantaneous flow divides m³/s and m³/h;
Velocity digit: The display of converter is 3-digit for choice according to the need;
Time: The velocity meter will sent average data within a certain period to display.This is
measurement period. Setting 0 means manual set measurement control time(Long
measurement period can improve stability of velocity display and signal output. It uses for
pulsating flow measurement field. Short measurement period means fast corresponding
speed uses for instant velocity measurement);
Interval: Interval between twice measuring 0

900s;

Value number: This parameter is for serial number which can be inputted by user. It uses
for recognizing value property. The value number auto add 1 every time;
Record: This parameter uses for historical data inquiry of value through page turning to
judge value property;
Auto shutdown: 1 is Time-limited model(1800s auto shutdown), 60s auto shutdown
without any operation. 0 is continuous working model;
Backlight: 0 is auto backlight shutdown when faulty operation. 1 is keep running.

7.2 Engineering mode menu

7.2.1 Engineering interface and parameters introduction

PAGE

/

-Model

Setting/Unit

-Model menu

Setting/Unit

Engineeri
ng
interface

Conversion
factor

Factory
calibration

Conversion factor

Real value

Instrument
factor

Factory
calibration

Zero calibration

Manually set the
zero point

Model

Factory
setting

Velocity cutoff

Small

velocity

cutoff

Level range

Factory
setting

Velocity digit

0

3

Max voltage

0

6.000

v

Transmission
address

1

255

Min voltage

0

1.600

v

Transmission
speed

1200

9600

Linear allow

0 or 1

Filter factor

10

200s

Velocity Value

1

5

Linear correction

0=off

1=on

Velocity
Calibration

1

5

Velocity Value

Factory setting

Sensor factor

0

6.0000

Sensor factor

Real value

Flow factor

0

6.0000

Language

0=Chinese

1=English

Password

0000

9999

Password

0000

Содержание MGG/KL-DCB

Страница 1: ...MGG KL DCB Portable Electromagnetic Velocity Meter Operation Manual...

Страница 2: ...measurement multiple parameters display self diagnosis alarming RS485 optional and storage function MGG KL DCB is the series model of our portable electromagnetic velocity flow meter This series incl...

Страница 3: ...parameter setting 5 7 1 Velocity meter model menu 6 7 2 Engineering mode menu 7 Instrument usage 8 8 1 Velocity measuring 8 8 2 Flow measuring 9 8 3 Measurement requirements 10 8 4 Operating mode 11 8...

Страница 4: ...ter conservancy II Main feature No movable parts no winding No choked flow no maintenance avoid troubles of other calibration and testing Micro power design can last 80 hours when replace the battery...

Страница 5: ...So we install a pair of electrodes on the pipe wall which orthogonal with sounding pipe axis and magnetic line The electrodes can detect the induced electromotive force It should be directly proporti...

Страница 6: ...ted coupler plug Please directly insert the plug and tighten the screws 6 1 2 Display interface Model Model display interface Model display interface 6 1 3 Battery and communication port The battery i...

Страница 7: ...4 6 2 Velocity sensor The velocity sensor consists of velocity sensor and tail cable Velocity Sensor The Tail...

Страница 8: ...assword input Under Engineering Mode The password must be inputted when modify status and parameter Password can be set arbitrarily in 5 digit number between 00000 99999 the default is 00000 In the wo...

Страница 9: ...s 1 m h Velocity digit 0 3 0 0 000 Time 1 10s Time 1 10s Interval 1 10s Interval 1 10s Value number 0 999 Reverse adjustment 0 off 1 on Record 0 999 Auto shutdown 0 off 1 on Auto shutdown 0 on 1 off...

Страница 10: ...uiry of value through page turning to judge value property Auto shutdown 1 is Time limited model 1800s auto shutdown 60s auto shutdown without any operation 0 is continuous working model Backlight 0 i...

Страница 11: ...rd It is adjustable parameter the factory default is 0000 Velocity cutoff Low velocity cutoff display unit is m s Model Transmission address The range of mailing address is 1 255 default is 1 Model Tr...

Страница 12: ...uring margin velocity measuring height side slope ratio channel bottom width water depth water level adjustment Model water level can be automatically collected by depth sensor Note The measuring meth...

Страница 13: ...0 2 V0 8 2 3 poin t 0 2 0 6 0 8h Vm 0 25 V0 2 2 V0 6 V0 8 4 poin t 0 2 0 4 0 7 0 9h Vm 0 25 V0 2 V0 4 V0 7 V0 9 5 poin t Close to surface 0 2 0 6 0 8h Riverbed Vm 0 1 V0 3V0 2 3V0 4 2V0 8 VS 6 poin t...

Страница 14: ...r parallels to the flow direction Waiting a moment for stable running of sensor then resend the data Avoid the primary interference signal input and eliminate the electrode oxide film interference 8 4...

Страница 15: ...calibration both medium and sensor 8 6 Historical data record and query Model display can record 1000 items historical data 8 6 1 Data record and storage The instrument will record internal storage nu...

Страница 16: ...1 Low press alarm Please replace battery 2 Exciting alarm 1 Excitation line loose 2 Excitation line fault 3 Return to factory 3 Bare electrode 1 Velocity measuring height setting is not current 2 Wat...

Страница 17: ...Dear Customer Thank you for your business Please complete the form below to improve our quality and service Your cooperation will be highly appreciate Product Model No Inspector Certification Problem...