6.2
Connecting the RTF Module to the iPA
Figure 5 Recommended orientation of RTF module when connected to iPA
USB Cable
Monitor Cable
1. Attach the RTF module to the iPA’s RF Output port and tighten the RF connector. Do not allow the body of the RTF
module to rotate while tightening the connector, as this can cause damage to the 7-16 connector mating surfaces.
The recommended orientation of the RTF module is shown in Figure 5.
2. Connect the AUX port of the RTF module to the iPA’s Monitor port using the cable provided.
3. Connect the USB port on the RTF module to the Type A USB port under the iPA’s protective side cover.
6.3
Calibrating the System
Before the RTF module can be used, it must first be calibrated according to the following procedure:
1.
If the iQA/iPA is currently turned off, turn it on now (refer to the relevant iQA/iPA Operating Manual for details
concerning the startup procedure).
2. There will be a brief delay while the iQA/iPA connects to the RTF module and searches for a valid calibration. For the
iQA, if a valid calibration is found, the following dialogue box is displayed:
Click “Ok” to complete the setup process. The instrument is now ready for use.