2.7 Additional equipment 

 

Nose opening  

If a nose opening is upholstered in the head part as an optional equipment feature, the opening can be closed 
with a cushion (optional). To open up the nose opening, reach under the head part when the table is standing 
still and press the cushion out from the bottom upwards (a slight amount of pressure is required). To close it, 
simply insert the cushion into the opening (a slight amount of pressure is required). 

 

Paper roll holder (optional equipment) 

The paper roll holder consists of a holding bar and angle mount brackets or retaining brackets. In addition to the 
stainless steel bar, the paper roll holding bar consists of a spring-loaded stainless steel sleeve featuring a round 
steel gripping disc at the end. To insert the holding bar, feed the guide cotter pin of the stainless steel bar into the 
rear hole of the angle mount bracket/retaining bracket. Then push the sleeve with the gripping disc  
against the spring force and feed the front guide cotter pin into the second hole. Then relieve the spring tension on 
the sleeve. To release the paper roll holder, proceed in the same way. 

 

3-sectioned head part (optional only with models 2890XLE, 2890XL/H, 2895XLE, 2895XL/H) 

The 3-sectioned head part consists of an adjustable head part as described under 2.3 and additional 
armrests, which can be lowered and removed steplessly, to the right and left of the head part. The clamping is 
released by loosening the knurled screw located underneath each armrest. The armrest can now be pulled 
down through a range of approx. 180 mm. Re-clamping takes place by subsequently tightening the knurled 
screw. To remove the armrests, loosen the knurled screw located directly underneath the head part. The 
complete armrest can then be pulled off the head part from the side. The armrests must not be used for 
sitting. They only serve to support the patient’s arms. 
 
 

2.8 Special features of models 2690XLE, 2695XLE 

The tables for echocardiography treatment have a cut-out on the left-hand side 
in the upholstered surface. It can be closed with a cushion. The cushion can be pushed out from the bottom 
upwards (a slight amount of pressure is required). To close the cut-out, first place the cushion in the rear of 
the cut-out and then press it down into the upholstered surface (a slight amount of pressure is required). The 
lower side of the cut-out is not parallel to the upper side. There is a visible slant. The position of the cushion in 
the cut-out is thus clearly identifiable and there is no risk of it being accidentally pushed out. 
In addition, model 2695XLE has a second cut-out on the right-hand side of the upholstered surface. 

 

Caution: Do not reach under the cushion or into the open cut-out during height adjustment. Always 
remove instruments/equipment from the cut-out area before adjusting the height. 

   

Potential equalisation connector

 

A potential equalisation connector according to DIN 42801 is attached to the long side of the underframe on 
the left (side with the upholstery cut-out). This connection must be used if interference from electromagnetic 
radiation can be seen on the monitors during the examination, which could falsify the measured values. 
Despite compliance with the applicable EMC regulations, the use of highly sensitive examination equipment 
can cause this minimal radiation to become visible. Only use connectors that comply with DIN. Through 
additional wiring of the individual assemblies, there is conductive contact between them. 

The potential equalisation connector is marked with the symbol

  

    

 
 

3. Additional accessories (for user-specific table configuration), in extracts 

 

 

 

Twin castors Ø 100 mm, centrally lockable  

Lowered underframe 

 

Paper roll holder  

 

Additional foot switch 

 

Additional hand switch   

 

Foot switch fixation on the underframe 

 

 

Side rails

 

 

Side guard for side rails

 

 

Содержание 2690XLE

Страница 1: ...setup and use 3 1 6 Commissioning 3 1 7 Safety instructions 3 1 8 Model designation and type labelling 4 1 9 Meaning of the serial number 4 2 Operating manual 5 2 1 Table design 5 2 2 Height adjustment 5 2 3 Head part adjustment 5 2 4 Middle part adjustment 5 2 5 Foot part adjustment 6 2 6 Movability model dependent or optional 6 2 7 Additional equipment 7 2 8 Special features of models 2690XLE 26...

Страница 2: ...sing the device We recommend that you keep these instructions for use near the device for future reference The device may only be used by authorised instructed and competent persons who are sufficiently familiar with its adjustment mechanism or have read and understood the operating manual fully The manufacturer cannot be held responsible for damage caused by or involving unauthorised persons No t...

Страница 3: ...e is ready for use upon delivery Remove the power cable from the film packaging on the underframe of the table and connect it properly to a permanently installed mains socket When routing the power cable make sure the cable cannot be crushed rolled over or otherwise damaged Lock the castors optional or the wheel system optional As the operator carry out a thorough and precise function check once t...

Страница 4: ...and type labelling The exact model designation depends on the choice of frame colour 00 white powder coated RAL 9010 03 white aluminium powder coated RAL 9006 04 grey aluminium powder coated RAL 9007 and the type of selected height adjustment E electromotive H hydraulic The type plate is attached to the underframe on one long side of the table It provides information about key table data The follo...

Страница 5: ... time of at least 400 s must be observed If the maximum duty cycle is exceeded an internal thermal switch protective temperature limiter in the motor interrupts the power supply to the actuator After the electric motor has cooled down the thermal switch automatically reconnects the power supply to the actuator Height adjustment with hydraulic height adjustment At tables equipped with hydraulic hei...

Страница 6: ...stors The castors can be locked using the foot operated brake on the castor housings In this case the castors can neither be moved nor rotated To release operate the brake on each castor again Please note Standard castors are not electrically conductive optional equipment You can recognise conductive castors via the respective marking yellow dot on the side of the running surface or a yellow ring ...

Страница 7: ... the head part from the side The armrests must not be used for sitting They only serve to support the patient s arms 2 8 Special features of models 2690XLE 2695XLE The tables for echocardiography treatment have a cut out on the left hand side in the upholstered surface It can be closed with a cushion The cushion can be pushed out from the bottom upwards a slight amount of pressure is required To c...

Страница 8: ...eight may vary depending on the equipment The motor is equipped with a thermal protective switch as standard This causes the motor to cut out if the weight load is too extreme or the motor s duty cycle DC 25 s 400 s 25 s 400 s is exceeded The table should be ready for use again after a 15 minute rest period There is therefore no risk of overloading the motor The mechanics of the tables have been d...

Страница 9: ...the separate enclosed sheet The table is equipped as standard with the Dewert hygiene standards which enable optimal cleaning and disinfection Hinge lids are made of the identical upholstery covering All undersides of the upholstery parts are coated with an upholstery covering and can thus be cleaned and disinfected Vent holes underneath the upholstery For the homogeneous foam to recover quickly a...

Страница 10: ... proper fit of the screw connections Lubricate the swivel joints and operating levers slightly with low viscosity spray oil if necessary Lubricate the roller guides with a little bearing grease or similar using a brush if necessary Insufficient lubrication is noticeable through noise development A checklist for maintenance technical inspection can be found in the appendix In addition to maintenanc...

Страница 11: ...p function is not observed the actuator cannot be operated After 30 seconds of operation in one direction of travel the actuator switches off and goes into sleep mode The actuator can still be operated for up to 3 seconds after the last operation to ensure fine adjustment Within this time window it is possible to move in each direction of travel again for a maximum of 30 seconds The actuator alway...

Страница 12: ...in the Dual System in accordance with the German Packaging Act VerpackG and therefore bear the disposal costs meaning you can dispose of the packaging free of charge Product Safety notices o Pay attention to sharp edges and pointed objects o When transporting the table only take hold of the underframe NOT the upper frame o In order to prevent accidents later on the no longer used product must be r...

Страница 13: ... purpose of curative and disease treatment examination massage and health therapy Table operation and patient positioning on the table may only be performed by professionally trained persons who have been instructed in its use or who through experience with other similar medical devices have knowledge of its proper use taking into account possible hazards The device must only be moved within the r...

Страница 14: ...d supply lines undamaged Function check With electromotive adjustment Move all motors with the foot switch or the hand switch to both limit positions until automatic switch off to ensure that the table operates smoothly without any collisions or blockages no cable connection can be overstretched crushed or otherwise damaged the motors run without noticeable noise development the end position switc...

Страница 15: ...le All necessary toggle screws present Possible movability Castors undamaged freewheeling given Connection to frame undamaged Re tighten all screw connections with central locking system also the grub screws of switch levers Safe braking effect Check with locked brakes by pulling and pushing the table Central Movability Levers must be in contact with the frame tighten the fastening screws Shift le...

Отзывы: