background image

Calibrating C4 & C6

Connect red hook to

test signal output

Leave black hook

un-connected

Insert hook probe to J1. Connect the red hook to the test signal

terminal J4 and leave the black hook un-connected.

1.

Set [SEN1] switch to 0.1V and [SEN2] switch to X5.

Set [CPL] switch to AC or DC.

2.

Adjust timebase to 0.2ms. You should see waveform similar

to that shown in photos below. If traces are not stable adjust

trigger level (the pink triangle on right screen border) so as

you get a stable display.

3.

Turn C4 (capacitor trimmer) with a small screw driver so that

the waveform displays sharp rightangle (photo C).

4.

C4

C6

Set [SEN1] switch to 1V and [SEN2] switch to X1while keep all other

settings unchanged. Adjust C6 so that sharp rightangle waveform is 

displayed.

5.

 Not enough

 Too much

 Good

Page 2

Test analog board

Step 3

1. Check voltages and controls

14. Hook Probes

Put wire through hood cap and sold the
wire onto hook terminal as shown.
Match wire color with hook color.

Finished look

Attach the main board to the 
analog board. Apply 5V DC 

power through J7.

1

2

3

Set switch [CPL] to GND, [SEN1]
to 1V, and [SEN2] to X5. 
Check voltages at the points as
shown in the photo.

VBUS
V+
AV+
V-
AV-
V1
V2
V3
V4

+5.10V
> 3.5V
> 3.5V

< -3.5V

< -3.5V

0V
1.05V
2.1 V
-1.05V

References

(*)

(*)

(*) Input dependent

V4

V2

V1

V3

AV+

V-

AV-

V+

Place negative

pen at GND

Couple switch

(set to GND)

Save Waveform

Press [SEL] & [+] buttons simultaneously. The currently displayed waveform

will be saved to EEPROM. The existing data in EEPROM will be over-written.

VPos Alignment

Measurements
ON/OFF

Functions

Operations

Move cursor to VPos indicator. Hold down [OK] for 3 seconds. Then follow screen prompts.

Default Restore

Recall Waveform

Center HPos

Center Trigger

Level

Move cursor to timebase. Hold down [OK] button for 3 seconds to turn ON or OFF on-screen

measurements including Vmax, Vmin, Vavr, Vpp, Vrms, Freq., Cycle, Pulse width, and Duty cycle.

Press [SEL] & [-] buttons simultaneously. 

Recalled waveform is always displayed

in Hold state.

Hold down [+] and [-] buttons simultaneously for about 3 seconds. 

Move cursor to the top bar. Hold down [OK] button for about 3 seconds. This will move
the display window to the center of capture buffer.

Move cursor to trigger level indicator. Hold down [OK] for 3 seconds. This will set
the trigger level to the medium value of signal amplitude.

Operations

[SEL] button: 

Select parameter to be adjusted. The selected parameter will be highlighted.

[+] and [-] button: 

Adjust the parameter selected by [SEL] button.

[OK] button: 

Freeze waveform refresh (entering HOLD state). Press on it again will de-freeze.

[CPL] switch: 

Set couple to DC, AC, or GND. When GND is selected the scope input is disconnected

from outside and connected to ground internally (0V input). 

[SEN1]/[SEN2] : 

Adjust sensitivity. The product of [SEN1] and [SEN2] settings makes the 

actual sensitivity which is displayed at the screen lower-left corner.

More Operations

Power supply voltage must not exceed
8V. 

Attention

1.

Allowed maximum signal input voltage
is 50Vpk (100Vpp) .

2.

Analog bandwidth
Sensitivity range

Resolution
Record length

Max realtime sample rate

Timebase range

Max input voltage
Input impedance

Power supply

Current consumption
Dimension

Weight

1MSa/s
0 -- 200KHz
10mV/div - 5V/div
50Vpk (1X probe)
1M ohm/20pF
12 bits
1024 points 
500s/Div -- 10us/Div

3.5V - 5V DC

< 100mA 

85 x 75 x 15 (mm)

50 gram (without probe)

Trigger modes
Trigger position range

Auto, Normal, and Single
Center (fixed)

Specifications

HOLD RUN

/

[OK]:

Parameter

Adjustment

[+] and [-]:

Parameter

Selection

[SEL]:

Trigger

State

Trigger Level

Indicator

Trigger

Slope

Trigger

Mode

Timebase

(s/div)

Sensitivity

(V/div)

Couple

Vertical

Position
Indicator

Power input

Couple

Selection

Sensitivity

Selection 1

Sensitivity

Selection 2

[CPL]:

[SEN1]:

[SEN2]:

Horizontal

Position

Oscilloscope

Mode

Signal Input

Display and Controls

4

Check slide switches and push-
buttons for correct operation.

(*)

(*)

Power input
(USB)

Reset (by shorting

the two pads)

5

Calibrate C4 & C6 if everything
is fine (see instructions to the

right).

Send Waveform

Data

Hold down [SEL] button for 3 seconds will send waveform data in texts via serial port

J5. The baudrate is 115200. Data format is 8N1.

Trigger sources
Ext. trigger thresholds
Ext. trigger input range

Internal/External
Low: 1.1V,  High: 2.2V

0V - 10V(max)

VBUS

V2 - V4 could have up to

 +/- 0.2V variance depending

on power supply voltage.  

Note:

Trigger

Source

Trigger Level

(internal only)

www.jyetech.com -

JYETech Ltd.

Tech Support: www.jyetech.com/forum

Toggle Test Signal

Amplitude

Move cursor to trigger slope indicator. Hold down [OK] button for 3 seconds to toggle test signal

amplitude between 3.3V and about 0.14V. The amplitude is indicated by         to screen top.

Calibrate Analog

Gain

Move cursor to trigger source indicator. Hold down [OK] button for 3 seconds to enter analog gain

calibration mode. Follow the on-screen instructions.

Ext. Trigger In

(PB15)

[H]

[J]

[Analog board 
ver. H shown]

[Analog board 
ver. H shown]

[Analog board 
ver. H shown]