background image

Advanced Operations

In order to use USB function the host which DSO112A is to communicate with is required

to install driver supporting the USB-Uart bridge CP2102. Please use the following link

to download driver and install it (refer to documents accompanying).

www.silabs.com/products/mcu/pages/usbtouartbridgevcpdrivers.aspx 

USB Connection

Serial Port Parameters

For the main firmware  serial port parameters are fixed to 115200 bps and 8-N-1.
For the bootloader  serial port parameters are fixed to 9600 bps and 8-N-1.

Waveform Data Upload

Please refer DSO112A User Manual for detailed instructions.

The DSO112A User Manual is available at 

Δ

Δ

T cursors and 

V cursors

Δ

 

T and 

cursors can be turn on or off by touching           or           button at the menu screen.

Δ

V

To change cursor position touch the cursor you want to change to highlight it.  Then touch

anywhere of the panel and scroll horizontally (for 

. You will see the

 highlighted cursor moves and readouts change accordingly (see Fig. 6 and Fig.78).

Δ

Δ

T) or vertically (for  V)

Maintenance

Charging the Battery

DSO112A contains a built-in smart Li-ion battery charger. Simply

connect the unit to USB power supply charging will start

automatically. A red LED will light up (which can be seen from

outside) indicating charge is undergoing. The charger will stop

automatically once the battery is fully charged. (Fig. 8)

Touch Panel Calibration

The touch panel has been calibrated before shipping. In case of the

touch panel doesn

t work properly it can be re-calibrated. Please

follow the steps below to perform touch panel calibration.

1 ) Power up the scope. At the main booting screen push [P]

button to enter Setup mode.

2 ) Press and hold [P] button in Setup mode to enter touch

panel calibration state, as shown in Fig. 9. 

3 ) Use a pointed object to touch the center of each cross 8 times,

as accurately as you can. You will see three crosses, one after another.  Do not touch more

than 8 times for each cross.

4 ) After you finish the third cross touch anywhere of the panel to verify. You should see a

small green cross follows the points you touch with some error (the error is usually

smaller than 1mm). If you see the errors larger than 2mm you can re-do the calibration.

5 ) Exit calibration by pressing [P] button.  

Upgrading Firmware

New firmware can be downloaded to DSO112A by the pre-installed Bootloader. Please follow the 

steps below to perform firmware change.

1 ) Download new firmware (it is  a file with 

.hex

 extension) to you computer. 

2 ) Download 

 at 

 and install it to

your computer. 

avrubd.exe

http://www.jyetech.com/Support/avrubd.rar

3 ) Connect DSO112A to PC via USB  and power up.
4 ) Start avrubd. At Option - System tab select 

Lite XModem

 as communication protocol

and set maximun connect count to 100 (see Fig. 10). 

5 ) At Option - Comport tab select the com port number corresponding to your scope and set

parameters to 9600 baudrate, 8 data bits, 1 stop bit, and no parity (see Fig. 11).

6 ) Click on the Open File icon to load 

the new firmware you have downloaded.

Click on the download icon to start downloading (see Fig. 12). You should see avrubd is

talking to oscilloscope by sending connection key continuously.

into avrubd 

7 ) At DSO112A enter Menu and touch the button              . Bootloader will be entered and 

answer to avrubd. If successful you should see download starts and progresses in avrubd.

After all data have been downloaded the scope re-starts and upgrading is done.

Specifications

Analog bandwidth: 0 

 2MHz

Max sampling rate: 2.5Msps

Sensitivity range: 2mV/Div 

 20V/Div

 

Sensitivity error: < 5%

 

Max input voltage: 50Vpk

Record length: 512/1024 points

Input impedance: 1M

Ω

 

(

.

)

Current consumption: ~300mA @3 7V

Highlighted
cursor

Highlighted
cursor

Charging LED

The unit of  T readout is the same as the current unit of timebase.

Δ

Fig. 6

Fig. 7

Fig. 8

Fig. 9

Fig. 10

Fig. 11

Fig. 12

Load HEX
file

Download

Note: 

The unit of  V readout is the same as the current unit of sensitivity.

Δ

Timebase range: 1 s/Div 

 50s/Div

μ

 

Dimension: 80 x 70 x 18 mm

 

www.jyetech.com

Presets

Preset Save/Load state is entered by touching               button at the menu. Touch a row to select 

(highlight) a  location. Touching              button will save current parameters to the selected location

(existing data will be over-written). Touching             will replace current parameters with that stored

in the selected location (if not empty). Touch 

Rename

 at top to change the title of the selected preset. 

External Trigger

External trigger input shares the same connector as test signal output (see Fig. 1). When trigger source

is set  to external the test signal will be disabled and trigger level will be displayed at top-right corner.

When trigger source is switched back to internal the test signal will resume. 

The allowed voltage range for external trigger signal is 0 - 15V.

www.jyetech.com/forum

Tech Support:

Отзывы: