background image

www.jyetech.com -

JYE Tech Ltd.

Page 3

2. Assemble Front Module

1. Solder Rotary Encoder

Note:

Please pay attention to the
orientation of PCB. Use the

side with outline marking.

Assembly Front Module

Step 3

Verify voltages on the analog board

Check Voltages

Step 4

Fit LCD to front panel as shown below.

1

2

Fold the main board over while keeping
LCD in place.

Final assembly 

Put all parts together

Step 6

3

Mount rotary encoder board to the
front plate with screws and solder the

board to J2 of the main board.

Mount with
screws
(KA2x4)

Make sure PCB evenly
touch the front plate
while doing C.

A.

B.

C.

Solder here

Calibration

Step 5

Screw the analog board to
back cover with the top 

bracket attached

1

Combine the front module and
the back cover

2

Screws
KA2x4

Top bracket

Put test signal 
terminal through
the small slot

Make sure receptacle (J4)
mate with pin-header (J2)

3

Attach bottom bracket before
holding the two modules

together

4

Bottom bracket

Attach the front frame

5

Firmly press the frame in

6

Front frame

Screw up at the back

7

Screws 
(4 positions)

PA2.3x8

Attach knob cap and done!

8

Knob cap

Attach the analog board to the main board
by mating J2 on the analog board to J4 on

the main board (see photo).

1

2

Apply 9V DC power supply to J7 (or J6) on
the main board.

3

Set couple switch to GND position.
Check voltages at the points
as shown in the photo.

4

V4

V2

V1

V3

AV+

V-

AV-

V+

Input
V+
AV+
V-
AV-
V1
V2
V3
V4

+9.30V
+8.35V

5.0V+/-2%

-5.0V+/-2%

-7.86V

0V
0V
0V
~1.65V

References

(*)

(*)

(*) Input dependent

Place negative

pen at DGND

Couple

switch

Apply power

here

Connect the red clip to the test signal terminal and leave the

black clip un-connected (see photo at bottom).

1

2

Apply power and boot. Hold down ADJ dial for 3 seconds to 

bring up Test Signal amplitude display at lower-left corner.

Push ADJ to set the amplitude to 0.1V.

3

Set sensitivity to 50mV and adjust trigger level so that

waveform stable (see 

How to Use

 at page 4).

Not enough

A.

Good

B.

Too much

C.

4

Tuning C3 so that sharp rectangle (photo B at left) is obtained.

The adjustment for C3 is done.

5

Similarly, for C5 calibration push ADJ to set test signal to 

3.3V. Change sensitivity to 1V. Tuning C5 so that sharp

rectangle waveform is obtained. The adjustment is done.

Adjust trimmers C3 and C5

Mount to the small PCB
(PN: 109-15002-00A)

(4 positions)

Important !

Always remove power before
connecting or disconnecting
the analog board.

Important !

Tips:

Perform VPos alignment
before calibrating C3 and C5.

Measurements applicable
for PCB version 
109-15001-00E or later

Analog board

part number

C3

C5

It is strongly recommended to read the article 

FAQ, Tips, and Troubleshooting

at 

 under the sub-forum 

DSO Shell

.

www.jyetech.com/forum

(DC jack is 5.5mm dia.

with 2.1mm core)

JYETech 

WWW.JYETECH.COM

Содержание DSO Shell 150

Страница 1: ... damaged 2 Identify IC orientation and diode polarity see photos 1 Before soldering check components against the part list to make sure you have correct parts Check kit contents and part quantities values by the photo at right and part list in page2 and page 3 Report missing or wrong parts to your vendor 1 Resistor values are easy to mis read It is strongly suggested to check their values by ohm m...

Страница 2: ... 9V power supply center positive to J7 on the board to check the display 1 2 You should see the scope boots up to a screen similar to the photo below D1 LED blinks twice Now apply power again Test power switch and tact buttons for their correct functions Apply power Check display 1 2 5 Pin header male 1X4 pin 0 1 pitch J2 Let iron stay on one pad of the resistor until solder on the other pad melt ...

Страница 3: ... Set couple switch to GND position Check voltages at the points as shown in the photo 4 V4 V2 V1 V3 AV V AV V Input V AV V AV V1 V2 V3 V4 9 30V 8 35V 5 0V 2 5 0V 2 7 86V 0V 0V 0V 1 65V References Input dependent Place negative pen at DGND Couple switch Apply power here Connect the red clip to the test signal terminal and leave the black clip un connected see photo at bottom 1 2 Apply power and boo...

Страница 4: ...nter Trigger Level Hold down OK button for about 3 seconds This will turn ON or OFF on screen display of measurements including Vmax Vmin Vavr Vpp Vrms Freq Cycle Pulse width and Duty cycle Press buttons simultaneously ADJ Trigger Recalled waveform is always displayed in Hold state Hold down SEC DIV and TRIGGER buttons simultaneously for about 3 seconds Hold down SEC DIV button for about 3 seconds...

Отзывы: