36
TROUBLESHOOTING
PROBLEM #1
Readout is blank and alarm "beeps"
- This indicates that incorrect or no information is
coming from the fish.
Possible causes:
1) Broken cable.
2) Damaged cable connector (where fish cable plugs into control box).
3) Bad preamp in fish.
PROBLEM #2
Proton 4 stops cycling
- One or both 12 volt batteries need recharged.
CALL FACTORY IF ANY PROBLEMS