Jungbrunnen Vitality Health Check Health Reader Скачать руководство пользователя страница 2

}

 

Vitality Health Check Health Reader – Quick Guide 

REF  VCH-R-001  

Jungbrunnen - FOY GmbH – Cantianstrasse 23 - D-10437 Berlin  

Quick Guide VHC-Reader- Status 12/17 

à

 Full Manual: www.vitality-health-check.com/service 

 

Error Codes 

 

INVALID  

Check if the test cassette is placed correctly below the cube and in the cassette adaptor.  

 

ERR

 

 

The device could not read the information from the RFID card. 

Press button briefly (<1 sec.), display will show 'ON' and try again.  

ERR XX

   

Battery error. Replace batteries.  

 

FAIL 

The device could not find a C-line. Check if the test cassette is placed correctly be-
low the cube and in the cassette adaptor. Check if C-line is clearly visible. If not, 
repeat the test. 

 

DATE

 

The expiry date of the test LOT seems to be exceeded. The device checks

 

the

 

inter-

nally set 

t

ime

 

with

 

the expiration date

 

of the test LOT.

´ 

Check the expiration

 

date 

of the test.

 

 

 
No Function  

Despite pressing the button no information is visible on the screen.  
The batteries may be discharged. 

Setting Date and Time 

 
Bring the device to position ‘ON’.  
Then, press the button shortly twice (<1 sec.). Year, date and time will appear on the display. 
 
Press the button for about 1 sec., a flashing display appears with the first-time specification, 
year. By repeated short (<1 sec.) pressing of the button, the displayed value can be changed. 
When reached the wanted value (e.g. year) press the button longer (>1 sec), the appointed 
value will be stored and the next time information will be presented.  
Repeat these steps to successively move to year, month, day, hour and minute.  
 
 
After setting the date and time information accordingly the device will display ‘OK’. 

 

Press the button one more time, the reader will show ‘ON’ and is now ready for use. Repeat 
this process after every battery change 

 
 

Attention:

  When turning on the VHC-Reader for the first time or after changing the batteries, date 

and time may need to be set.  

 

Отзывы: