12
Installation
Details
•
in flush-mounted box
•
Installation note UPS module: installation into deep flush box
Installation height in accordance with DIN VDE 0834
(call systems in hospitals, nursing homes and similar facilities)
•
Pull cord push-button
min. installation height: at least 20 cm
above the shower head
•
Operating devices
0.7 to 1.5 m above the floor
•
Operating devices with text
1.5 to 1.7 m above the floor
•
Devices with installation units
1.6 to 1.8 m above the floor
•
Signal lights and text displays
1.5 to 2.2 m above the floor
DIN 18024-2 “Accessibility” also prescribes to install operational elements for wheelchair
users at a height of 0.85 m.
Installation and connection of calling systems is only permitted for trained electricians.
Installation
Installation above the supporting frame without claws
According to DIN VDE 0834 wearing components, i.e. batteries and accumulators, must
be replaced once a year.
Disconnect plug connector before changing the accumulator or decommissioning the
emergency system. This will prevent the accumulator from discharging and short-circuits
during the decommissioning of the accumulator.