Screen Display

Operating Instructions

Displayed State

Sound and

Indicator Color

The display is as shown , when

there is no more data checked while

recalling memories. Remove 2 dry

batteries and after 10 seconds

re-install the power to clear all

memory data.

With a long beep,

the green indicator

light shows up and

turn red, when

device is powered

on.

Switching between mute and un-mute

In the power-on state, press and

hold the sound switch button for

about 2 seconds to switch the sound

on or off. When the sound is turned

on, it will beep once and the mute

symbol will be displayed when the

sound is off.

The

symbol is displayed in

Mute mode and disappears in

Un-mute mode.

When the sound is

turned on, the green

indicator light

shows up with a

long beep.

Switching between object temperature and body temperature

Содержание PD-FR409-B

Страница 1: ...JPD FR409 BT Infrared Thermometer Instruction Manual Shenzhen Jumper Medical Equipment Co Ltd Manual Version 1 0 Date of Issue 2019 8 0482...

Страница 2: ...without the prior consent of Shenzhen Jumper Medical Equipment Co Ltd All statements information and recommendations in this document are provided AS IS without warranties guarantees or representatio...

Страница 3: ...Display and Operating Instructions 11 Measuring Ear Temperature 18 Measuring Forehead Temperature 20 Measuring Object Temperature 21 Connecting the Thermometer to an APP Application of a Mobile Phone...

Страница 4: ...accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment do...

Страница 5: ...l need the following information when making your claim device model serial number purchase date and your contact information and address Package Contents No Name Quantity 1 Infrared Thermometer 1 2 P...

Страница 6: ...o not use the thermometer on newborns or for continuous temperature monitoring purposes Do not use the thermometer for purposes that are not specified in this User s Manual Follow the instructions in...

Страница 7: ...end the overdue thermometer to local garbage disposal site or send it back to us 2 AAA batteries of 1 5V are the only replaceable accessories of the thermometer Please do not use the batteries of othe...

Страница 8: ...e thermometer may be damaged permanently Do not take temperature measurements on body parts other than forehead and ears Otherwise the temperature readings may be inaccurate During measurement do not...

Страница 9: ...y or damage to the thermometer may occur if the thermometer is not used correctly Attention Inaccurate reading or damage to the thermometer may occur if the thermometer is not used correctly Non ioniz...

Страница 10: ...d adults have higher body temperatures than the elderly Women s body temperatures are approximately 0 3 C 0 5 F higher than men s Variation in body temperature Body Site Normal Temperature Range Foreh...

Страница 11: ...frared temperature sensor and a Microprocessor 3 Operating principle The infrared temperature sensor collects infrared energy emitted by the eardrum or the skin surface After being focused by a lens t...

Страница 12: ...red temperature sensor with high sensitivity High accuracy with automatic temperature calibration 05 Diverse functions 20 readings recall Fever alert Switching between C and F Automatic power off powe...

Страница 13: ...measuring the ear temperature 7 Unit switch button 8 Probe cover put the cover on when measuring the forehead temperature 9 Memory button Sound switch 10 Battery cover Display description 1 Object te...

Страница 14: ...etween 35 0 C 95 0 and 37 5 C 99 5 there will be a long beep and a green indicator This indicates that your body temperature is normal When the forehead temperature is between 37 6 C 99 6 and 42 2 C 1...

Страница 15: ...e Mode button the thermometer enters the Ear mode The symbol Ear is displayed on the screen Switch to the adult or child measurement mode by switching the side adult child button according to your mea...

Страница 16: ...hermometer enters the Forehead mode The Head symbol is displayed on the screen Switch to the adult or child measurement mode by switching the side adult child button according to your measurement need...

Страница 17: ...e screen See the table in the Sounds and indicator color instructions section Out of the measuring range display In Ear mode a temperature reading of more than 42 0 C 107 6 F In Forehead mode a temper...

Страница 18: ...shown followed by the recorded reading Press the Memory button again for the next recorded data 02 will be shown followed by the recorded reading A maximum of 20 temperature readings can be recalled W...

Страница 19: ...ws up and turn red when device is powered on Switching between mute and un mute In the power on state press and hold the sound switch button for about 2 seconds to switch the sound on or off When the...

Страница 20: ...d body temperature Body temperature contains the Forehead temperature and Ear temperature The silent green indicator light Switching between F C In the power off state remove the battery cover and tog...

Страница 21: ...ndicator light An error occurs when data is being read from or written to the memory or the temperature correction is not complete A long beep with the red indicator light When the battery voltage is...

Страница 22: ...reen 5 When measuring an adult switch the button to the adult mode When measuring children switch the button to the child mode and the symbol is displayed on the screen 6 Insert the temperature probe...

Страница 23: ...ld whose ear canal is small Measuring Forehead Temperature 1 Put the cover on the probe of thermometer 2 Press the Power button to power on the thermometer 3 Press the Mode button the thermometer ente...

Страница 24: ...ode The symbol is displayed on the screen 4 Point the thermometer probe to the center of the object about 1 3cm away from the object surface 5 Press and release the Measure button for 1 second The tem...

Страница 25: ...ise the temperature probe may be polluted and temperature readings may be inaccurate 3 Do not hold the thermometer for a long time because it is sensitive to the ambient temperature 4 Make sure the se...

Страница 26: ...ery cover off along the marked direction and take it off 2 Insert the two AAA batteries into the compartment according to the stated polarities Make sure that the batteries are installed correctly Oth...

Страница 27: ...water out off the lens during the cleaning process Otherwise the lens may be damaged The lens may be scratched if it is cleaned with a hard object which might result in inaccurate readings Do not clea...

Страница 28: ...reventive inspection maintenance period 1 Ensure the safety of thermometer and check whether it has potential safety hazards in normal use each week e g whether the lens is broken the shell has cracks...

Страница 29: ...e between 10 C 50 0 F and 40 C 104 F The temperature reading is lower than the typical body temperature range The lens of the temperature probe is dirty Clean the lens using a cotton swab The thermome...

Страница 30: ...de 35 0 C 42 0 C 95 0 F 107 6 F Object mode 0 0 C 100 0 C 32 0 F 212 0 F Accuracy Laboratory Forehead mode 0 2 C 0 4 F Ear mode 0 2 C 0 4 F Object mode 1 0 2 0 Resolution 0 1 C 0 1 F Memory 20 tempera...

Страница 31: ...ess of water IP22 Safety degree of using in flammable anesthetic gas blending with air oxygen or nitrous oxide Non AP APG No application parts of the thermometer prevents defibrillation charge effect...

Страница 32: ...nce and manufacturer s declaration Electromagnetic emission for all equipment and systems Guidance and manufacturer s declaration Electromagnetic emission The Infrared Thermometer JPD FR409 BT is inte...

Страница 33: ...ation Electromagnetic immunity The Infrared Thermometer JPD FR409 BT is intended for use in the electromagnetic environment specified below The customer or the user of the Infrared Thermometer JPD FR4...

Страница 34: ...R409 BT is intended for use in the electromagnetic environment specified below The customer or the user of the Infrared Thermometer JPD FR409 BT should assure that it is used in such an environment Im...

Страница 35: ...ansmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an Electromagnetic a site survey a should be less than the c...

Страница 36: ...vel above the JPD FR409 BT should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the JPD FR409 BT b...

Страница 37: ...d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter m...

Страница 38: ...sed failures 1 Failure resulting from unauthorized disassembly and modification 2 Failure resulting from an unexpected dropping during application or transportation 3 Failure resulting from not follow...

Страница 39: ......

Страница 40: ...Shenzhen Jumper Medical Equipment Co Ltd Address D Building No 71 Xintian Road Fuyong Street Baoan Shenzhen Guangdong China Tel 86 755 26696279 Fax 86 755 26852025 Website http www jumper medical com...

Отзывы: