29
function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The Infrared Thermometer JPD-FR409 is
suitable for use in all establishments,
including domestic establishments and
those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic
purposes.
Guidance and manufacturer's declaration – Electromagnetic
immunity –for all equipment and systems
Guidance and manufacturer´s declaration – Electromagnetic immunity
The Infrared Thermometer JPD-FR409 is intended for use in the
electromagnetic environment specified below. The customer or the user of the
Infrared Thermometer JPD-FR409 should assure that it is used in such an
environment.
Immunity test
IEC
60601
test level
Compliance
level
Electromagnetic
environment- guidance
Electrostatic
discharge
(ESD)
±6kV
contact
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with