Model : JPD-300E
Fetal Monitor
Instruction Manual
Version: V1.0
Compile date: Sep. 2017
Shenzhen Jumper Medical Equipment Co., Ltd.
Страница 1: ...Model JPD 300E Fetal Monitor Instruction Manual Version V1 0 Compile date Sep 2017 Shenzhen Jumper Medical Equipment Co Ltd...
Страница 2: ...of the product 1 Use this product in accordance with the Instruction Manuel Technical Specifications 2 The installation maintenance and upgrade of the product shall be carried out by the personnel of...
Страница 3: ...me Shenzhen Jumper Medical Equipment Co Ltd Address D Building No 71 Xintian Road Fuyong Street Baoan Shenzhen Guangdong China Tel 0755 26696279 26692303 Fax 0755 26852025 Postal Code 518103 Website H...
Страница 4: ...onfirm that the fetus is the source of the FHR that you are monitoring It is to be understood that MHR traces can present features that are very similar to FHR traces and even include acceleration and...
Страница 5: ...7 4 2 Connect the power supply Charge 18 5 Operational Guidelines 18 5 1 Preparation before Use 18 5 2 FHR monitoring 18 5 3 TOCO monitoring 19 5 4 Manual FM monitoring 19 5 5 Monitoring information u...
Страница 6: ...n accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment...
Страница 7: ...ards The whole machine design is simple reliable and easy to use The main features of this product are as follows The portable Fetal Monitor can be used in the hospital or at home The host and the pro...
Страница 8: ...cy is manually recorded by fetal movement pen button according to the pregnant women feel their own fetal movement Monitoring all stages after pregnancy Fetal Monitor is not applicable to monitor wate...
Страница 9: ...ck whether the repeated use of accessories are damaged whether the sensor is damaged whether cable connections are reliable replace if necessary and properly handle the damaged accessories according t...
Страница 10: ...ggest that under the premise of meeting the clinical needs the ultrasound irradiation time to pregnant women should be as short as possible Caution Do not use instrument immediately when it is transfe...
Страница 11: ...tant Caution The accuracy of the fetal heart rate is controlled by the device and cannot be adjusted by the user If FHR result is not accurate please use other methods For example use a stethoscope to...
Страница 12: ...ction and other functions can be conducted through the shuttle keys including system parameter settings freeze display volume control and alarm control and so on Ultrasound fetal heart probe US Ultras...
Страница 13: ...ce 2 In the power on state press switch button and hold it for 2 seconds to turn off the device 3 In power on state each short press the switch button to update the monitoring 3 2 3 Fetal movement pen...
Страница 14: ...the selection function The cursor will move when shuttle button rotates Short press the shuttle button to confirm if enter the submenu or select a parameter 3 2 5 Uterine contraction back to zero but...
Страница 15: ...e metal contact of the uterine contraction probe touch the sub contact of the host charging terminal The probe will automatically pair with the host all power indicator lights are on indicating the pa...
Страница 16: ...through rotating to the left and increasing sound through rotating to the right Data the state of the data that is currently connected to the central station The icon is displayed when the connection...
Страница 17: ...connected with the ultrasonic fetal heart probe When the host is paired with the ultrasonic fetal heart probe and the probe is on the white background of the label will be on and the current electric...
Страница 18: ...limit lower FHR limit warning switch etc Host Version View the current version of the host Probe version view the current version of ultrasonic FHR probe Return Return to the main interface Reset Pas...
Страница 19: ...3 4 2 WiFi settings 1 Access to the WiFi configuration interface automatically search the nearby WiFi rotate the shuttle button to select the hotspot to connect enter the password and select OK to co...
Страница 20: ...setting static IP interface then could set fixed IP subnet mask and default gateway 3 IP and port number in the configuration process are IP address and port number connected with the central station...
Страница 21: ...election pressing the button to confirm exit by pressing Back to set the number of the device the settable range is 00 to 99 3 4 4 FHR Warning Figure 10 FHR Warning Interface FHR Warning selects on wh...
Страница 22: ...set password enter the new password the new password allows the maximum number of characters to be entered in 10 bits click OK after the completion of entry to complete password reset 4 Setting of mon...
Страница 23: ...toring interface remove the ultrasonic probe the TOCO probe and the FM pen from the host if the FHR probe connection identifier and TOCO probe connection identifier of the screen light up it indicates...
Страница 24: ...the strap adjust the position the tightness of the strap should be appropriate 5 3 2 Press the uterine contraction pressure reset button on the host at which point the uterine contraction position sho...
Страница 25: ...e and other strong solvents Caution Do not use wearable material such as steel wool or silver polishing agent Caution Do not allow any liquid to enter the enclosure or immerse any part of the instrume...
Страница 26: ...humidity range 95 c Atmospheric pressure range 50kPa 106kPa 7 3 Storage Packed monitors must be stored in well ventilated room with temperature of 20 55 relative humidity no more than 80 and without c...
Страница 27: ...WiFi function and configure WiFi Check the router The screen displays the probe icon in gray Unpaired The probe is charging Retake the probe and align to the contact point for matching Normal phenome...
Страница 28: ...y other language without the prior written consent of our company The content in this manual may be subject to change without further notice Warning The intended use of this device is clinical and can...
Страница 29: ...0 240V 50 60Hz internal battery lithium battery3 7V 3000mAh Input Power 20VA Work mode continuous working for more than 10 hours after full power supply Size 89 2mm length 202mm width 75 9mm height Ne...
Страница 30: ...verage sound intensity of space peak time shall not exceed 100mW cm2 The sound intensity of the device type shall not exceed 5mW cm2 Gel viscous water based compound is not irritative and allergic to...
Страница 31: ...of Aaprt X 1 2 1 2 Y 1 2 1 2 OthersInf ormation td 903 9 prr 1250 pr at maximum Ipi 0 018 deq at maximum Ipi 1 2 Ipa 3 at maximum MI 0 00 Operation ControlC ondition Control 1 Control 2 Control 3 Note...
Страница 32: ...re advice Note Before using this device please make sure that all of the requirements for EMC in this section are met Note This Chapter shall list the contents described in IEC 60601 1 2 and the users...
Страница 33: ...eclaration electromagnetic immunity The JPD 300Eis intended for use in the electromagnetic environment specified below The customer or the user of the JPD 300E should assure that it is used in such an...
Страница 34: ...4 8 3 A m 3 A m Power frequency magnetic fields should be at levels haracteristic of a typical location in a typical commercial or hospital environment Note UT means AC network voltage before applied...
Страница 35: ...Hz 10 V m Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM For EQUIPMENT and SYSTEMS that are not LIFE SUPPORTING Recommended separa...
Страница 36: ...ated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 M...
Страница 37: ...mper Medical Equipment Co Ltd Address D Building No 71 Xintian Road Fuyong Street Baoan Shenzhen Guangdong China Tel 86 755 26692192 26696279 Fax 86 755 26852025 Website www jumper medical com Postal...