5 Function check

When the paperless recorder is installed and connected, it can be started up.
After the supply voltage has been connected or switched on, the start screen
will appear briefly.

Start screen

After the end of the initialization phase, the visualization level is automatically
started.

Visualization
level

The recorder is now in recording mode. 

Further steps

A

The instrument can be configured by an authorized person, either
from the instrument keys or by using the setup program. Additional
information about the configuration can be found in the Operating
Manual B 95.5011.0.

Afterwards, please check again that connection, configuration and
operational mode are correct. 

Содержание LOGOSCREEN Series

Страница 1: ...Paperless Recorder for secure acquisition of FDA compliant measurement data B 95 5011 4 Installation Instructions 11 04 00415650 es...

Страница 2: ...Menu structure of the paperless recorder...

Страница 3: ...ameplate 9 2 2 Type designation 10 2 3 Standard accessories 11 2 4 Accessories 11 3 Installation 13 3 1 Location and climatic conditions 13 3 2 Mounting in position 13 4 Electrical connection 15 4 1 I...

Страница 4: ...Contents...

Страница 5: ...commissioning please do not carry out any manipulations on the unit as this could endanger your rights under the instrument warranty Please contact the nearest subsidiary or the head office in such a...

Страница 6: ...an introduction to the system and not as a formal technical documentation Instrument documentation in printed form B 95 5011 1 Operating Instructions The operating instructions are an extract from the...

Страница 7: ...n on the connection and use of modules of the JUMO mTRON automation system B 95 5011 2 3 Interface description PROFIBUS DP interface This provides information on the connection of a paperless recorder...

Страница 8: ...used where special precautions must be taken when handling electrostatically sensitive components Note signs H Note This symbol is used to draw your special attention to a remark v Reference This sig...

Страница 9: ...r The production number indicates the production date year week The figures concerned are in position 12 13 14 15 Example F Nr 0022969000003130006 This shows that the paperless recorder was manufactur...

Страница 10: ...x x x 68 RS232C and PROFIBUS DP x x x x x x 69 RS422 485 and PROFIBUS DP 5 Internal memory x x x x x x 0032 32MB backup memory x x x x x x 0064 64MB backup memory x x x x x x 0128 128MB backup memory...

Страница 11: ...ef of the sensor connection cables 1 CD with additional documentation see Chapter 1 2 Arrangement of the documentation 2 4 Accessories PC software package consisting of setup program PC evaluation sof...

Страница 12: ...2 Identifying the instrument version 12...

Страница 13: ...possible from shock and vibration Stray electromagnetic fields from motors transformers etc should be avoided The ambient temperature at the location can be 0 to 45 C at a relative hu midity of 75 no...

Страница 14: ...the front into the panel cut out h From the back of the panel hook the two mounting brackets into the recesses on the sides of the housing The flat sides of the brackets must be against the housing h...

Страница 15: ...d screened cables Earth the screen on the instrument side to the earth potential k Earth the instrument at the PE terminal to the protective conductor This cable should have the same cross section as...

Страница 16: ...onnection 16 4 2 Procedure h Carry out electrical connection as per Chapter 4 3 Connection diagram h Screw back panel cover on spacer bolt first h If necessary use cable ties for strain relief of conn...

Страница 17: ...panel Rear view with pluggable screw terminals Cover for back panel V The electrical connection must only be carried out by qualified personnel L L Hole for lead sealing Holes provided for the cable t...

Страница 18: ...ce thermometer in 4 wire circuit 1 to 12 Resistance transmitter 1 to 12 Potentiometer in 2 wire circuit 1 to 12 Potentiometer in 3 wire circuit 1 to 12 Potentiometer in 4 wire circuit 1 to 12 Voltage...

Страница 19: ...ial 6 VP supply voltage Plus 8 RxD TxD N A cable receive transmit data N 21 Ethernet RJ45 socket extra code 1 TX transmit data 2 TX transmit data 3 RX receive data 6 RX receive data 22 Setup interface...

Страница 20: ...3 to 5V DC HIGH 12 to 30V DC 1 24V 50mA auxiliary supply not stabilized 2 GND 5 Logic input 7 6 Logic input 6 7 Logic input 5 8 Logic input 4 9 Logic input 3 10 Logic input 2 11 Logic input 1 32 Exam...

Страница 21: ...n phase the visualization level is automatically started Visualization level The recorder is now in recording mode Further steps A The instrument can be configured by an authorized person either from...

Страница 22: ...5 Function check 22...

Страница 23: ...counter 1 ________________________________________________ Ext counter 2 ________________________________________________ 5 Read off IP address Subnet Mask and Gateway on the device or through the se...

Страница 24: ...th the password that was initially assigned by the administrator If a user no longer knows the password then the user status has to be set to New and a new password assigned through the PCS software A...

Страница 25: ...l 50 to 100 C adjustable through setup software Sampling cycle 6 or 12 channels 125msec Input filter 2nd order digital filter filter constant adjustable from 0 to 10 0sec Test voltage 500V across opto...

Страница 26: ...rrent up to 180 up to 390 up to 2000 up to 4000 150m 300m 2 4 500 A 250 A 500 A 250 A Connection type resistance transmitter 3 wire circuit potentiometer 2 3 wire circuit Shortest span 6 Sensor lead r...

Страница 27: ...ected Potentiometer not detected detected Voltage up to 1V not detected detected Voltage 1V not detected not detected Current not detected not detected 1 programmable reaction of instrument e g trigge...

Страница 28: ...ltage error 0 1 of range span Power consumption approx 25VA Electrical connection at rear by plug in screw terminals max conductor cross section 2 5mm2 or 2x 1 5mm2 with ferrules Ambient temperature r...

Страница 29: ...gement of 6 E Electrical connection 15 Electrostatic discharge ESD 5 F Fitting into the panel 14 Function check 21 I Identification of namplate 9 Installation 13 Installation notes 15 Instrument docum...

Страница 30: ...Relay outputs 19 Repeat order for recorder in the identical version 9 Returning 5 S Sales documentation in printed form 6 Standard accessories 11 Start screen 21 Supply 18 T Technical data 25 Type de...

Страница 31: ......

Страница 32: ...6003 607 e mail mail jumo net Internet www jumo net JUMO Instrument Co Ltd JUMO House Temple Bank Riverway Harlow Essex CM20 2TT UK Phone 44 1279 635533 Fax 44 1279 635262 e mail sales jumo co uk Inte...

Отзывы: