29
10 Configuration
10.8
Digital inputs of base unit and optional boards
Open:
Device Menu > Configuration > Digital Inputs >
Digital Input 1 to 9
Response at hold
low
high
NAMUR low
NAMUR high
hold
safety value
Specification of the analog output value when the
hold function is activated, during calibration of
one of the sensors for the particular output or at
error (overrange/underrange)
low:
lower limit of the standard signal value
range
[
0 V or 0(4) mA
]
high:
upper limit of the standard signal value
range (10 V or 20 mA)
NAMUR low:
lower NAMUR limit of the standard
signal
[
0 V or 0(3.4) mA
]
NAMUR high:
upper NAMUR limit of the stan-
dard signal (10.7 V or 22 mA)
hold:
unchanging analog value
Safety value:
see configuration item
"Safety value" in this table
Response during cali-
bration
Moving
Frozen
Safe value
Response in case of a
fault
low
high
NAMUR low
NAMUR high
hold
safety value
a
The unit for the value set for the "Signal output value" appears in the entry
field.
Configuration point Selection/
setting option
Explanation
Configuration point Selection/
setting option
Explanation
Description
up to 21 text characters
Designation for the input
Inversion
Yes
No
Invert/do not invert state
Contact
Base unit:
potential-free contact,
external voltage source
Type of digital signal connected
Optional boards:
potential-free contact
Alarm
Binary inputs alarms are used to monitor input-side switching signals. The
alarm settings for all digital device functions are explained together.
⇨
Содержание AQUIS touch S
Страница 8: ...Inhalt ...
Страница 20: ...10 Configuration 20 Chapter 10 12 Calibration timer page 46 ...
Страница 68: ...13 Calibrating redox sensors 68 ...
Страница 78: ...15 Calibrating CR conductivity sensors 78 ...
Страница 88: ...16 Calibrating Ci conductivity sensors 88 ...
Страница 96: ...17 Calibrating universal inputs 96 ...
Страница 106: ...18 Calibrating O DO sensors 106 ...
Страница 146: ...21 Annex 146 ...
Страница 147: ......
Страница 148: ......
Страница 149: ......