3 Installation specifications for immersion sleeve

3

Installation specifications for immersion sleeve

NOTE!

Mounting in thermowell only with plastic screw connection.

1. Pinch together the plastic screw connection on the probe cable.

Important:

 Do 

not

 mount the plastic screw connection into the recesses (beading) on the tempera-

ture probe sleeve.

2. Push the plastic screw connection up to the probe tube. Screw the probe into the immersion sleeve.

3. Tighten the screw connection with a tightening torque of 3 to 5 Nm (hand-tight).

Make sure that the probe is mounted securely in the immersion sleeve (probe tip must sit on the base
of the immersion sleeve).

(3)

(1)

(2)

(3)

Содержание 902442

Страница 1: ...90244261T94Z001K000 V2 00 EN 00669425 Screw connection for temperature probe installation in plastic or brass Installation Instructions...

Страница 2: ......

Страница 3: ...tents Contents 1 Safety information 4 2 Installation specifications for direct mounting 5 2 1 Plastic screw connection 5 2 2 Brass screw connection 6 3 Installation specifications for immersion sleeve...

Страница 4: ...accessible to all users at all times If difficulties occur during startup please do not intervene in any way that could jeopardize your warranty rights Warning symbols CAUTION Risk of burns The instal...

Страница 5: ...in the assembly aid to determine the probe installation length Grasp the two halves of the plastic screw directly above the assembly aid and press into the 3 re cesses beads on the temperature probe...

Страница 6: ...a loosely placed slotted pin 2 Insert the back of the temperature probe into the assembly aid up to the limit stop 28 mm 3 Press in the slotted pin until it is flush with the hexagonal surface e g usi...

Страница 7: ...hrough the O ring and into the measuring point and tighten the screw connection tightening torque 3 to 5 Nm when installed correctly the bottom of the hexagon sits on top of the standard compliant mea...

Страница 8: ...able Important Do not mount the plastic screw connection into the recesses beading on the tempera ture probe sleeve 2 Push the plastic screw connection up to the probe tube Screw the probe into the im...

Страница 9: ......

Страница 10: ...ers Road East Syracuse NY 13057 USA Delivery address Mackenrodtstra e 14 36039 Fulda Germany Phone Fax Email Internet 44 1279 63 55 33 44 1279 62 50 29 sales jumo co uk www jumo co uk Phone Fax Email...

Отзывы: