10 Operation
46
10.8 Calibration level
All the calibrations that have been enabled by the administrator (administrator
level) can be carried out in this level.
✱
Press the
key for at least 3 seconds.
✱
Use the
or
key to select CALIBRATION LEVEL.
10.8.1 REL. CELL CONSTANT (relative cell constant)
If this function has been enabled by the administrator, then the operator can
calibrate the relative cell constant of the device here;
see "Calibrating the relative cell constant", Page 51.
10.8.2 TEMPCO LINEAR (linear temperature coefficient)
If this function has been enabled by the administrator, then the operator can
calibrate the device for liquids with a linear temperature coefficient;
see "Linear temperature coefficient (ALPHA)", Page 52.
10.8.3 TEMPCO NON-LIN. (non-linear temperature coefficient)
If this function has been enabled by the administrator, then the operator can
calibrate the device for liquids with a non-linear temperature coefficient;
see "Non-linear temperature coefficient (ALPHA)", Page 55.