73
25 Configuration level of the universal indicator
25.3 Configuration parameter for general functions
(not specific to the universal indicator)
Chapter 26 “Configuration level (not instrument-specific)”, page 74.
25.4 nuLL - SLoP - SiL - SiH
SLoP
Slope correction (end value)
1
nuLL
Zero point correction (offset)
1
SiL
Start of transmission range
This value is taken from the operating instructions for the attached instrument.
Example for a JUMO 202630
(transmitter for free chlorine -> transmission range = 0 — 2.0 mg/l):
SiL = 0.00
SiH
End of transmission range
This value is taken from the operating instructions for the attached instrument.
Example for a JUMO 202630
(transmitter for free chlorine -> transmission range = 0 — 2.0 mg/l):
SiL = 2.00
1
This parameter does
not
provide any indication of the true state of the
probe that is connected.