Preparations
22
Installation of the magnetic valve
JULABO Order-No. 8 980 704 3
Magnetic valve
Fix the expansion tank (43) to the connection (41) (open
end wrench SW19)
Screw the magnetic valve to the entry for cooling water
(45). Respect the flow direction (arrow)
Connect the cooling water to the magnetic valve and to the
outflow of cooling water (44) by means of tubes 8 mm ID.
Fix the control cable (56) at the control exit (56a) and srew
tightly.
Adapter to reduce the pump pressure
JULABO Order-No.
8 970 802
With this adapter the pump pressure at the pump connection
(51) can be reduced from 1,2 bar to
0,8 bar
–0,1
.
Assembly
Fix the adapter to reduce the pump pressure to the pump
connections (50, 51) (open end wrench SW19).
Notice:
When using pressure sensitive glass vessels, it is absolutely necessary to work with
an adapter to reduce the pump pressure.
Caution
:
Securely attach all tubing to prevent slipping.
43
45
41