![Julabo HL-4 Скачать руководство пользователя страница 82](http://html1.mh-extra.com/html/julabo/hl-4/hl-4_operating-manual_2030693082.webp)
Remote control
82
Important times for a command transmission:
To ensure secure data transfer, the time gap between two commands
should be at least 250 ms.
The circulator automatically responds to an
in
command with a data string
followed by an LF (Line Feed). The next command should be sent only
after 10 ms.
The
out
commands are valid only in remote control mode.
Examples:
When the RS485 interface is used, the three-digit instrument address
precedes each command.
(example: address Ad32 = A
0 3 2
)
Command to set the working temperature >Setpoint1< to 55.5 °C
out_sp_00
55.5
A032_out_sp_00
55.5
Command to retrieve the working temperature >Setpoint1<
in_sp_00
A032_in_sp_00
Response from the circulator:
55.5
LF
A032_55.5
LF
Notice:
out commands
Temperature values can be displayed on the circulator in °C or °F.
The settings sent via the interface must also be made in the selected unit according
to this setting.
12.3. List of commands
out commands:
Setting parameters or temperature values.
Command
Parameter
Response of circulator
out_mode_01
0
Use working temperature >Setpoint1<
out_mode_01
1
Use working temperature >Setpoint2<
out_mode_01
2
Use working temperature >Setpoint3<
out_mode_02
0
Selftuning “off”.
Temperature control using the stored parameters.
out_mode_02 1
Selftuning
“once”
Single selftuning of the controlled system after the next start.
out_mode_02 2
Selftuning
“always”
Continual selftuning of the controlled system whenever a new
setpoint is to be reached.