Menu functions
42
ACTUAL
SET
Ad¯P
-
Calibration of programmer input:
Highest value = 80 °C
ACTUAL
SET
Ad
_
P
-
Calibration of programmer input:
Lowest value = -20 °C
ACTUAL
SET
Calibration procedure:
c
In the configuration level, set the programmer
input type to allow control via an external
programmer.
0 = Voltage
0 to 10 V
1 = Current
0 to 24 mA
d
Connect an external programmer to the connector
(O).
e
Set the external programmer to the highest
temperature value.
ACTUAL
SET
f
In the calibration level, select the item Ad¯P and
set the highest temperature value
(example: 80 °C).
n
Press the setpoint key (E).
Follow the instructions under section 6.3.1.
page 26
o
p
q
g
Do not alter any settings on either of the units as
long as the display "Ad¯P" is blinking.