![Julabo 9400330 Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/julabo/9400330/9400330_operating-manual_2030695027.webp)
Cryo-Compact Circulator
27
Warning:
For supervised or unsupervised operation with the AUTOSTART function, avoid
any hazardous situation to persons or property.
The Cryo-compact circulator does no longer conform to N.A.M.U.R.
recommendations.
6.7.
Remote control: activate – deactivate
(Interface OFF)
(Interface On)
Ð
Ð
Ð
The Cryo-Compact Circulator is to be prepared for remote control
by a personal computer via the serial interface RS232.
Set the interface item
from >IOFF<
(Interface OFF)
to
>ION<
(Interface On).
Remote control: activate – deactivate:
x
Switch off the Cryo-Compact Circulator by pressing the mains
switch and wait approx. 5 seconds.
x
Keep depressed the keys
and enter
simultaneously
and turn on the unit with the mains power switch.
>I OFF<
No remote control via RS232 (Factory setting)
>I On<
Remote control via RS232
L
The software version and the type of unit is indicated
(see example on the left).
The display "
r OFF
" indicates the unit is ready to be operated
via remote control.
www.
GlobalTestSupply
.com
Find Quality Products Online at: