background image

©JRI  

g)

 

 Manual start  

Press shortly on the pushbutton  

 
 

 
 

 
 

 
 

h)

 

Alarm visualisation 

 
 

 
 

i)

 

Measurement stop 

Depending on the configuration, the SPY RF can stop recording or not. The different options are: 

Rolling memory: once the memory is full, the new values replace the old ones.  

Full memory: the recorder stops when its memory is full. 
With the software: you can put the SPY RF in standby mode with Sirius when you do not use your recorder. 

 

j)

 

Leds and pushbutton actions functioning  

The green led is on 2’’ when the measurement starts and then flashes each 1’ in recording mode. 

Specials functioning regarding the recorder using mode : 
 

Device set up in storage mode 

Pushbutton pressing 

Mode  

 < 5" 

5’’> pressing <10’’ 

OFF   

The 2 leds are on and flash 

at the same time. 

Starting measurements  

  

  

 

Pushbutton

  

Green led 2" = beginning of 

measurements 

 

 

Delayed (date & time) 

  

 

 

 

Immediately 

 

- - 

Measuring 

Green led 10" = auto control 

 

Alarm LED (red): 

Flashes every 15’’. 

Working LED (green): 
2’’ => starting measurements 

then flashes every 1 minute  

Содержание Spy RF AIR

Страница 1: ...NOTICE D UTILISATION Spy RF AIR 10077...

Страница 2: ...TALLATION 4 a Installation 4 V USE 5 a Stop 5 b Start 5 c Waiting mode 5 d Configuration 5 e Measurement start 5 f Automatic start 5 g Manual start 6 h Alarm visualisation 6 i Measurement stop 6 j Led...

Страница 3: ...ubbish dump or in a domestic waste container Comply to the regulation to throw away the device CE MARKING this equipment is certified to comply with the European regulation for the electric security i...

Страница 4: ...JRI 4 III PRESENTATION IV INSTALLATION a Installation Bouton poussoir T moin d alarme T moin de fonctionnement Capots ajour s pour la ventilation des capteurs 55mm...

Страница 5: ...configuration or to start a new recording session The symbol Halt is on no measures in progress Use the pushbutton to start d Configuration SPY RF configuration is done from the Sirius software and t...

Страница 6: ...ius when you do not use your recorder j Leds and pushbutton actions functioning The green led is on 2 when the measurement starts and then flashes each 1 in recording mode Specials functioning regardi...

Страница 7: ...ide this range Factory gauge inaccuracy 1 2 Operating conditions 30 C 70 C Temperature for storage 40 C 85 C Memory size 1s to 90 min Recording interval 20 000 measurements Radio range in free field 1...

Страница 8: ...d by specialists That s why JRI has created maintenance contracts We bring you a global solution which makes your maintenance easier This overall service offer includes maintenance and also metrologic...

Отзывы: