7.4 Selecting a program for operation
To select a stored program, press over SEL (Fig 7.4).
(SEL indication changes its color to yellow).
Once selected, this program in placed on the TASK screen.
7.5 Operation screen
On the operation screen «TASK», the program selected (PROG 04) on the
programming screen is shown (Fig 7.5) and it starts when pressing over .
This screen shows:
Actual temperature
Temperature set
Working time set (SET)
Elapsed time
Fig 7.4. Selecting a program
PROGRAM 04
SAVE
TIME: 00:00
TEMP: 50.0ºC
OVERTEMP: 0ºC
UNDERTEMP: 0ºC
SETUP
HH:MM:SS DD/MM/YY
SEL
SEL change to yellow
colour when program is
selected
Fig 7.5. Operation screen
HH:MM:SS DD/MM/YY
28.9
ºC
SET
50.0
FLOW
80%
PROG 04
SET -- h -- m
T
00 h 00m
7.6 Circulation pump
The liquid circulation pump could operate in two configuration modes:
-
INTERNAL circulation
:
Water (or other) will be pumped from horizontal pump output.
(See fig 7.6.1).
Use this configuration to make water circulation into the bath vessel in order
to homogenize the temperature from sample load.
-
EXTERNAL circulation
:
Water (or other) will be pumped from vertical pump output (See fig 7.6.2).
Use this configuration to made water circulation out of the bath vessel in order
to temperate an external equipment. This is the factory setting.
Fig 7.6.1
INTERNAL circulation
.
Fig 7.6.2
EXTERNAL circulation
.
Fig 7.3 screen is displayed.
Select a program (from 00 to 09) with keys:
Set working temperature and the other parameters:
Press:
Keys
will be shown to select temperature.
In a similar way press:
Sets de working time.
A time setting value of 00.00 means «conti-
nuous operation».
Sets the undertemperature alarm value.
Sets the overtemperature alarm value.
Press to save .
UNDERTMP: 0ºC
OVERTMP: 0ºC
SAVE
TEMP: 16.0ºC
TIME: 00:00
PROGRAM 00
SAVE
TIME: 00:00
TEMPERATURE: 16.0ºC
OVERTEMP: 0ºC
UNDERTEMP: 0ºC
SETUP
HH:MM:SS DD/MM/YY
Fig 7.3. Programming screen
Selected program.
Program selector.
SEL