
* The performance of the device may be degraded should one or more of the following occur:
-
Operation outside the manufacturer s s tated te mperature a nd h umidity r ange.
’
-
Storage outside the manufacturer s s tated te mperature a nd h umidity r ange.
’
-
Mechanical shock (for example, drop test) or degraded sensor..
-
Patient temperature is below ambient temperature.
* Portable and mobile RF communications can affect the device. The device needs special pre-cautions
regarding EMC according to the EMC information provided in the accompany documents.
124mm
PLEASE READ CAREFULLY BEFORE USING
CONTENTS
OWNER S M
ANUAL
’
1 Thermometer, 1 Owner s Manual, 1 Storage Case
’
Model:DMT-4735b
DIGITAL THERMOMETER
PRECAUTION
SPECIFICATIONS
Figure 1
SYMBOL EXPLANATION
℃
/
℉
SWITCHABLE
Warning :
Read instructions thoroughly before using digital thermometer.
Choking Hazard: Thermometer cap and battery may be fatal if swallowed. Do not allow children
to use this device without parental supervision.
Do not use thermometer in ear. Designed use is for oral, rectal, and armpit (axilla) readings only.
Do not place thermometer battery near extreme heat as it may explode.
Note: Use of the probe cover may result in a 0.01
℃
(0.02
℉
) discrepancy from actual temperature.
Remove battery from the device when not in operation for a long time.
The use of temperature readings for self-diagnosis is dangerous. Consult your doctor for the
interpretation of results. Self-diagnosis may lead to the worsening of existing disease conditions.
Do not attempt measurements when the thermometer is wet as inaccurate readings may result.
Do not bite the thermometer. Doing so may lead to breakage and/or injury.
Do not attempt to disassemble or repair the thermometer. Doing so may result in inaccurate readings.
After each use, disinfect the thermometer especially in case the device is used by more than one person.
Do not force the thermometer into the rectum. Stop insertion and abort the measurement when pain is
present. Failure to do so may lead to injury.
Do not use thermometer orally after being used rectally.
For children who are two years old or younger, please do not use the devices orally.
If the unit has been stored at temperatures over 5
℃~
40
℃
(41
℉~
104
℉)
, leave it in 5
℃~
40
℃
(41
℉~
104
℉)
ambient temperature for about 15 minutes before using it.
±
0.05
℃
(
±
0.1
℉
) during 35.50
℃
~42.00
℃
(95.90
℉
~107.60
℉
)
±
0.1
℃
(
±
0.2
℉
) during T<35.50
℃(
95.90
℉)
or T>42.00
℃
(107.60
℉
)
One 3.0V DC button battery type CR2032
Temperature readings are available in Celsius or Fahrenheit(located in the upper right corner of LCD.).
With the device off, press and hold the On/Off Button for approx. 3 seconds to change the unit and when the
thermometer connected with the APP ,it will synchronize the current unit. The LCD display will briefly shows
full segments at current unit mode with beep stone. Then the display shows another unitmode after change.
Followed by last stored temperature , the thermometer is now in the testing mode.
TROUBLESHOOTING
Problem
Solution
Error message
The system is not
functioning properly.
Unload the battery, wait for 1 minute and repower
it. If the message reappears, contact the retailer
for service.
Temperature taken is
lower than 32.00°C(89.60°F)
Turn off, wait one minute and take a new temperature
via close contact and sufficient rest.
Temperature taken is higher
than 43.99°C(111.18°F)
Dead battery: Battery icon is
flashing, can’t be measurable. Suggest to replace the battery.
Turn off, wait one minute and take a new temperature
via close contact and sufficient rest.
LIMITED WARRANTY
The thermometer is guaranteed for one year from the date of purchase. If the thermometer does not
function properly due to defective components or poor workmanship, we will repair or replace it free of
charge. All components are covered by this warranty excluding the battery. The warranty does not cover
damages to your thermometer due to improper handling.
To obtain warranty service, an original or copy
of the sales receipt from the original retailer is required.
Disposal of this product and used batteries should be carried out
in accordance with the national regulations forthe disposal of
electronic products.
Document No. JDMT-5604-017
Article No.: 005
Version: Z
Date of Issue: 2018.06.09
Measure Range:
Ambient operating range:
Storage and transportation
condition:
Liquid crystal display, 4 1/2 digits
For storing the last measured value
Eight month with 3 measurements per day
13.4cm
×
3.1cm
×
1.6cm (L x W x H)
Approx. 26 grams including battery
Temperature:
(
)
5
℃~
40
℃
41
℉~
104
℉
Relative humidity: 15%~95%RH
Atmospheric Pressure : 700hPa ~ 1060hPa
Temperature:
(
)
-20
℃
~55
℃
-4
℉~
131
℉
Relative humidity: 15%~95%RH
Atmospheric Pressure : 700hPa ~ 1060hPa
Classification:
Type:
Digital Thermometer (Not Predictive)
Type BF
Direct Mode
Ingress Protection Rating:
IP 27
Accuracy:
Operating mode:
Display:
Memory:
Battery:
Battery life:
Dimension:
Weight:
Expected service life:
Three years
32.00
℃-
43.99
℃
(89.60
℉
-111.18
℉
)(
℃
/
℉
chosen by manufacturer)
3
9
0
.1
7
PRODUCT FUNCTIONS & FEATURES
This product has two parts functions : the first function can be used as a basic thermometer for monitoring
ovulation cycles and other physiological information, this function should be used by women to measure
basal body temperature. the second function can be used as a normal thermometer ,used to measure and record
body temperature data, this function suitable for most people to use .
HOW TO USE THIS PRODUCT
1.Please download and install APP before using this product. Then register new account use your email account
and select your purpose, ovulation monitoring or normal thermometer.
2.Turn on the phone bluetooth through APP and press the on/off button next to LCD display to turn on the device.
The display will briefly shows full segment at current unit mode with beep tone. Followed by last stored
temperature , and shows 37.00
℃
or
98.60
℉.
Then select the device ID by the APP and bind it. While it
connected successful , APP will be synchronized with the thermometer.
*Note:
①
. If the measured temperature is less than 32
.
00
℃
or 89
.
60
℉
, the LCD will display “Lo” below
the graduation lines without scale bar, if the measured temperature is higher than 43
.
9
9
℃
or 111.18
℉
, the LCD will display all scale bars with “Hi” above.
②.
When
the LCD display “----” , it means that the thermometer is waiting for Bluetooth connection, you
only have 30 seconds to connect your thermometer , otherwise it will skip the Bluetooth connection into
the normal temperature measurement. Or you can press the on/off button to skip the waiting for Bluetooth
connection into the normal temperature immediately.
TAKE YOUR TEMPERATURE
PART1: For normal thermometer , position thermometer in desired location (mouth, rectum, or armpit.)
a)
Oral Use:
Place thermometer under tongue as indicated by
“√”
position shown in Figure 2. Close your mouth and breathe evenly
through the nose to prevent the measurement from being influenced
by inhaled/exhaled air. Normal temperature between 35.70
℃
and 37.30
℃
(96.26
℉
and 99.14
℉
)
b)
Rectal Use:
Lubricate silver probe tip with petroleum jelly for easy insertion. Gently insert
sensor approximately 1cm (less than 1/2 ") into rectum. Normal temperature between 36.20
℃
and 37.70
℃
(97.16
℉
and 99.86
℉
).
c)
Armpit Use:
Wipe armpit dry. Place probe in armpit and keep arm pressed firmly at side.
From a medical viewpoint, this method will always provide inaccurate readings, and should not
be used if precise measurements are required. Normal temperature between 35.20
℃
and 36.70
℃
(95.36
℉
and 98.06
℉
).
PART2: For basal body temperature ,while wake up in the morning ,open the APP and measurement method
reference to above Oral Use
WARM TIPS
①
.
When the peak temperature has been reached, the
℃
or
℉
will stop flashing and the thermometer will sound a series of beeps.
The minimum measurement time until the signaling tone (beep) must be maintained without exception. The measurement
continues even after the buzzer notification. So that in order to achieve better body temperature measurement result,
recommend to keep the probe in mouth and rectum about 2 minutes, or in armpit about 5 minutes regardless of the beep sound
and at least 30 seconds measurement interval should be maintained.
②
. To prolong battery life, press the
On/Off
Button to turn unit off after testing is complete. If no action is taken, the unit will
automatically shut off after around10 minutes.
PRODUCT ILLUSTRATION
On/Off
Button
LCD
Probe
Figure 1
BATTERY REPLACEMENT
1. Replace battery when
“
”
appears in the lower right corner of LCD display.
2. Put a thin board such as a coin on fillister of cover. Turn the battery anti-clockwise until the cover is off
(See Figure 3).
3. Use a non-meteal instrument such as a pen to remove old battery from the battery holder (See Figure 4).
Discard battery according to local law.
4. Place a new
into the chamber with positive side facing up.(See Figure 5).
5. With a thin pin to turn the cover clockwise until the
“
”
facing towards
“
”
(See Figure 6).
3.0V DC CR2032
Figure 4
Figure 3
Figure 5
Figure 6
Figure 2
JOYTECH HEALTHCARE CO.LTD.
No.365, Wuzhou Road, Yuhang Economic Development Zone, Hangzhou city
, 311100 Zhejiang,China Tel:+86-571-81957767
CLEANING AND DISINFECTION
Wipe the thermometer with a soft clean cloth.For stubborn stains, wipe the thermometer with
a cloth that has been dampened with water or a neutral detergent solution and then wring
thoroughly. Finish by wiping with a soft dry cloth.For disinfection, 75% Ethanol or Isopropyl
alcohol can be used.Observe the following to prevent damage to the thermometer.
-Do not use benzene, thinner, gasoline or other strong solvents to clean the thermometer.
-Do not attempt to disinfect the sensing section (tip) of the thermometer by immersing in
alcohol or in hot water (water over 5
0℃
(122
℉
))
.
-Do not use ultrasonic washing to clean the thermometer.
CALIBRATION
The thermometer is initially calibrated at the time of manufacture. If the thermometer is used according to
the use instruction, periodic readjustment is not required. However, we recommend checking calibration
every two years or whenever clinical accuracy of the thermometer is in question. Turn on the thermometer
and insert into the water bath and then check the laboratory accuracy of thermometer. Please send the
complete device to the dealers or manufacturer.
The above recommendations do not supersede the legal requirements. The user must always comply with
legal requirements for the control of the measurement, functionality, and accuracy of the device which
are required by the scope of relevant laws, directives or ordinances where the device is used.
Direct Current
Type BF Applied Part
Batch Code
Manufacturer
Consult Accompanying Documents
Storage and Transportation
Temperature Limit: -20
℃
~55
℃
(-4
℉ ~
131
℉)
-20
℃
55
℃
The first num.2:Protercted against access to hazardous parts with a finger,and the jointed test
finger of 12mm
φ,
80
, 80 mm length, shall have adequate clearance from hazardous parts .
And protected against solid foreign objects of 12,5 mm Ф and greater. The second number 7:
Protected against water and the test
is
made
by
completely
immersing
the
enclosure
in
water
between 0.15m and 1m for about 30 minutes.
The
water
temperature
does
not
differ
from
that
of
the
equipment
by
more
than
5
K.
IP 27
This digital thermometer provides a quick and highly accurate reading of an individual's body
temperature. The digital thermometer is intended to measure the human body’s temperature in
regular mode orally, rectally or under the arm, and the device is reusable for clinical or home
use on people of all ages. To better understand its functions and to provide years of dependable
results, please read all instructions first.
This appliance conforms to the following standards:
ASTM E1112 Standard Specification for Electronic Thermometer for Intermittent Determination of
Patient Temperature,
ISO 80601-2-56 Medical electrical equipment —Part 2-56:Particular requirements for basic safety
and essential performance of clinical thermometers for body temperature measurement,
IEC 60601-1-11 Medical electrical equipment —Part 1-11: General requirements for basic safety and
essential performance –Collateral Standard: Requirements for medical electrical equipment and
medical electrical systems used in the home healthcare environment and complies with the
requirements of IEC 60601-1-2(EMC), AAMI/ANSI ES60601-1(Safety) standards. And the
manufacturer is ISO 13485 certified.
FCC INFORMATION
Caution: Changes or modifications to this unit not expressly approved by the party responsible for compliance
could void the user authority to operate the equipment. This
device
c
omplies
w
ith
P
art
1
5
o
f
t
he
F
CC
R
ules
.
O
peration
i
s
s
ubject
t
o
t
he
f
ollowing
t
wo
c
onditions
:
(
1
)
t
his
d
evice
m
ay
n
ot
c
ause
h
armful
i
nterference
,
a
nd
(
2
)
t
his
d
evice
m
ust
a
ccept
a
ny
i
nterference
r
eceived
,
i
ncluding
i
nterference
t
hat
m
ay
c
ause
u
ndesired
o
peration
.
*Note:
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment generates, uses, and can radiate radio
frequency energy. If this equipment does cause harmful interference to radio or television reception,
which can be determined by turning the equipment off and on, the user is encouraged to try and correct
the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the distance between the equipment and the receiver.
- Connect the equipment to an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
Made in China