Care And Cleaning
1. The probe window must be kept clean, dry, and undamaged at all
times to ensure accurate readings. The accuracy of temperature
readings can be affected by damage to the probe window, or the
presence of dirt, fingerprints, earwax, dust and other soiling
compounds on the probe window. Degraded sensors can degrade
performance or cause other problems.
2
.
For cleaning
:
1) Soak a clean soft cloth in drinking water,wring it out,and then
wipe the thermometer(including probe) no less than 3 times
;
2) Visual or use magnifying glass to observe the thermometer have
no visible dirt and stains ,then use another clean soft cloth to
wipe the thermometer residue water
;
3) Put the thermometer in the original packaging.
3.For disinfection
:
1) Soak a clean soft cloth in drinking water,wring it out,and then
wipe the thermometer(including probe) no less than 3 times
;
2) Visual or use magnifying glass to observe the thermometer have
no visible dirt and stains ,then use another clean soft cloth to
wipe the thermometer residue water
;
3) Using a clean soft cloth dipped in 70% medical alcohol,wipe the
probe for 3 Times,each time 1 minute.
4) Using a clean cotton swab dipped in 70% medical alcohol,wipe
the sensor window 3 times
;
5) Wait at least 10 minutes to let the alcohol volatilization and put it
in the original packaging.
4. Do not put the thermometer into water directly.
5. Store the thermometer in a dry location, free from dust and
contamination and away from direct sunlight.
6.
Put the thermometer back to the original packaging after using.
Battery Replacement
1. Replace battery when " appears in the
right corner
"
upper
of LCD display. (See Figure )
7
2. Slide battery cover down as shown in Figure
8.
3. Remove battery and install 2 new AAA alkaline batteries
as shown
.
in Figure
9
4. Slide battery cover back on.
Figure
7
Figure
8
Figure
9
Specifications
Forehead mode: 34.0
℃
~43.0
℃
(93.2
℉
~109.4
℉
)
Object mode: 0
℃
~100
℃
(32
℉
~212
℉
)
Forehead mode:
±
0.2
℃
(0.4
℉
) during 35.5
℃
~42.0
℃
(95.9
℉
~107.6
℉
)
at 15
℃
~35
℃
(59.0
℉
~95.0
℉
) operating temperature range
±
0.3
℃
(0.5
℉
) for other measuring and operating
temperature range
Object mode:
±
4% or
±
2
℃
(4
℉
) whichever is greater
5
41
℃
~40
℃
(
℉
~104
℉
), 15%~85%RH, non-condensing
70kPa~ 106kPa
Atmospheric Pressure :
155*39.2*49.4mm
Approx. grams(with batteries)
97
DC3V(2
×
AAA
battery)
Measuring range
Laboratory accuracy
Display resolution
0.1
℃
or 0.1
℉
Measure time
Approximately 1 second
Operating temperature
range:
Storage and transport
temperature range
-25
℃
~ 55
℃
(-13
℉
~131
℉
), 15%~95%RH, non-condensing
70kPa~ 106kPa
Atmospheric Pressure :
Clinical accuracy
Dimension
Weight
Battery
Battery life
Approx. 1 year/3000 readings
Measuring site
Forehead(Forehead Mode)
Oral (This thermometer converts the forehead temperature
to display its “oral equivalent.”)
Reference body site
Operation mode Forehead mode(Adjust mode)
Ingress protecting rating
IP22
Shock
withstands drop of 3 feet
Expected service life
Three years
Clinical bias: -0.09
℃(-
0.16
℉)
Clinical repeatability: 0.13
℃(
0.23
℉)
Limits of agreement: 0.87
℃(
1.57
℉)
Contraindication
No contraindication
Troubleshooting
Problem
Solution
Error message
The thermometer is not
functioning properly.
Unload the battery,
wait for 1 minute and
repower it. If the
message reappears,
contact the retailer
for service.
The ambient temperature
is not within the range
between
℃
and 40
℃
(
℉
~
104
℉
).
5
41
Place the
in a room for at least 30
minutes at room
temperature between
℃
and 40
℃
(
℉
~
104
℉
)
thermometer
5
41