Calibration
The thermometer is initially calibrated at the time of manufacture.
If the thermometer is used according to the use instruction, periodic
readjustment is not required. However, We recommends checking
calibration every two years or whenever clinical accuracy of the
thermometer is in question. Please send the complete device to the
dealers or manufacturer.
The above recommendations do not supersede the legal requirements.
The user must always comply with legal requirements for the control
of the measurement, functionality, and accuracy of the device which
are required by the scope of relevant laws, directives or ordinances
where the device is used.
A clinical summary and procedures for checking calibration are
available upon request.(Turn on the thermometer and press the power
button long time until entering into calibrate mode, software
version
will be displayed.)
Direct Current
Disposal of this product and used batteries should be carried out in accordance
with the national regulations for the disposal of electronic products
Type B F applied part
Consult Accompanying Documents
LOT
Batch Code
26
Troubleshooting
Troubleshooting
In Ear mode:
Temperature taken is
higher than 43.0
℃
(109.4
℉
).
In Object mode:
Temperature taken is
higher than 100
℃
(212
℉
).
Problem
Problem
Solution
Solution
The ambient temperature
is not within the range
between 10
℃
and 40
℃
(50
℉
~
104
℉
).
Place the
in a room for at least 30
minutes at room
temperature between
10
℃
and 40
℃
(50
℉
~
104
℉
)
thermometer
Read Temperature
Taking Hints thoroughly ,
then make sure the probe
cover and lens filter are
clean, then take a new
temperature measurement.
The thermometer is
placed incorrectly
or unsteady.
The thermometer could
not
work due to low
battery.
The thermometer
showing a rapid ambient
temperature change.
The thermometer is not
functioning properly.
Read Illustration For
Use thoroughly and
take a new temperature
measurement.
Replace two new
alkaline batteries
size AAA.
Allow the thermometer
to rest in a room for at
least 30 minutes at
room temperature:
between 10
℃
and 40
℃
(50
℉
~
104
℉
)
Unload the battery,
wait for 1 minute and
repower it. If the
message reappears,
contact the retailer
for service.
Error message
Error message
Take a measurement
until or appears
on the display.
Read Temperature Taking
Hints Thoroughly, then
check the integrity of the
probe cover and take a new
temperature measurement.
24
Measurement before
thermometer is ready
Remove probe cover in
object temperature mode.
25
In object temperature
mode, measurement
with probe cover.
In Ear mode:
Temperature taken is
lower than 32.0
℃
(89.6
℉
).
In Object mode:
Temperature taken is
lower than 0
℃
(32
℉
).
23
Storage and Transportation Temperature Limit:
-13
℉℃~
131
℉
(-25
℃
~55
℃)
-25
℃
55
℃
Specifications
Ear/Object mode:
10
℃
~40
℃
(50
℉
~104
℉
), 15%~85%RH, non-condensing
00hPa ~ 1060hPa
Atmospheric Pressure : 7
46*36*143mm
Approx.83grams(with batteries)
DC3V(2
×
AA
battery)
Measuring range
Laboratory accuracy
Display resolution
0.1
℃
or 0.1
℉
Measure time
Approximately one second
Operating temperature
range:
Storage and transport
temperature range
-25
℃
~ 55
℃
(-13
℉
~131
℉
), 15%~95%RH, non-condensing
00hPa ~ 1060hPa
Atmospheric Pressure : 7
Clinical accuracy
Dimension
Weight
Battery
Battery life
Approx. 1 year/6000 readings
Measuring site
Ear canal(Ear Mode)
Oral (This thermometer converts the ear temperature to
display its “oral equivalent.”)
Reference body site
Operation mode Ear mode(Adjust mode)
Ingress protecting rating
IP22
Clinical bias
:
0.08
℃(
0.14
℉)
Clinical repeatability
:
0.13
℃(
0.23
℉)
Limits of agreement:0.73
℃(
1.31
℉)
Shock
withstands drop of 3 feet
Expected service life
Three years
Ear mode: 32.0
℃
~43.0
℃
(89.6
℉
~109.4
℉
)
Object mode: 0
℃
~100
℃
(32
℉
~212
℉
)
Ear mode:
±
0.2
℃
(0.4
℉
) during 35.5
℃
~42.0
℃
(95.9
℉
~107.6
℉
)
at 15
℃
~35
℃
(59.0
℉
~95.0
℉
) operating temperature range
±
0.3
℃
(0.5
℉
) for other measuring and operating
temperature range
Object mode:
±
4% or
±
2
℃
(4
℉
) whichever is greater
14