background image

Calibration

The thermometer is initially calibrated at the time of manufacture. 
If the thermometer is used according to the use instruction, periodic 
readjustment is not required. However, We recommends checking 
calibration every two years or whenever clinical accuracy of the 
thermometer is in question. Please send the complete device to the 
dealers or manufacturer. 

The above recommendations do not supersede the legal requirements. 
The user must always comply with legal requirements for the control 
of the measurement, functionality, and accuracy of the device which 
are required by the scope of relevant laws, directives or ordinances 
where the device is used.

A clinical summary and procedures for checking calibration are 
available upon request.(Turn on the thermometer and press the power
button long time until entering into calibrate mode, software

 

version

will be displayed.)

 

Direct Current

 Disposal of this product and used batteries should be carried out in accordance 
 with the national regulations for the disposal of electronic products

Type B F applied part

 

 

 

         
    

Consult Accompanying Documents

LOT

Batch Code

26

Troubleshooting

Troubleshooting

In Ear mode:
Temperature taken is 
higher than 43.0 

(109.4

).

In Object mode:
Temperature taken is 
higher than 100 

(212

).

Problem

Problem

Solution

Solution

The ambient temperature
 is not within the range 
between 10

 and 40

(50

~

104

).

Place the 
 in a room for at least 30
 minutes at room 
temperature between 
10

 and 40

(50

~

104

)

 thermometer

Read Temperature 
Taking Hints thoroughly ,
then make sure the probe 
cover and lens filter are 
clean, then  take a new 
temperature measurement. 

The  thermometer is 
placed incorrectly
or unsteady. 

The  thermometer could 
not

 

work due to low 

battery.

The  thermometer
 showing a rapid ambient 
temperature change.  

The  thermometer is not 
functioning properly.

Read Illustration For 
Use thoroughly and 
take a new temperature
 measurement.
 

Replace two new 
alkaline batteries 
size AAA.
 

Allow the thermometer 
to rest in a room for at 
least 30 minutes at 
room temperature: 
between 10

 and 40

(50

~

104

)

 Unload the battery, 

wait for 1 minute and 
repower it. If the 
message reappears, 
contact the retailer 
for service.
 

Error message              

Error message              

Take a measurement 
until      or     appears 
on the display.

Read Temperature Taking 
Hints Thoroughly, then 
check the integrity of the 
probe cover and take a new 
temperature measurement. 

24

Measurement before 
thermometer is ready

Remove probe cover in 
object temperature mode.

25

In object temperature
mode, measurement
with probe cover.

In Ear mode:
Temperature taken is 
lower than 32.0 

(89.6

).

In Object mode:
Temperature taken is 
lower than 0 

(32

).

23

Storage and Transportation Temperature Limit: 
-13

℉℃~

131

(-25

~55

℃)

-25

55

Specifications

Ear/Object mode:
10

~40

(50

~104

), 15%~85%RH, non-condensing

00hPa ~ 1060hPa

Atmospheric Pressure : 7

46*36*143mm

Approx.83grams(with batteries)

DC3V(2

×

AA

 

battery)

Measuring range

Laboratory accuracy

Display resolution

0.1

 or 0.1

Measure time

Approximately one second

Operating temperature 
range:

Storage and transport 
temperature range

-25

~ 55

 (-13

~131

), 15%~95%RH, non-condensing

00hPa ~ 1060hPa

Atmospheric Pressure : 7

Clinical accuracy

Dimension

Weight

Battery

Battery life

Approx. 1 year/6000 readings 

Measuring site    

        

Ear canal(Ear Mode)

 Oral (This thermometer converts the ear temperature to 
 display its “oral equivalent.”)

Reference body site

Operation mode                  Ear mode(Adjust mode)

Ingress protecting rating

         
    

 IP22

Clinical bias

0.08

℃(

0.14

℉)

Clinical repeatability

0.13

℃(

0.23

℉)

Limits of agreement:0.73

℃(

1.31

℉)

Shock

withstands drop of 3 feet

Expected service life

Three years 

Ear mode: 32.0

~43.0

(89.6

~109.4

)

Object mode: 0

~100

(32

~212

)

Ear mode: 

±

0.2

 (0.4

) during 35.5

~42.0

 (95.9

~107.6

at 15

~35

 (59.0

~95.0

) operating temperature range

±

0.3

 (0.5

) for other measuring and operating 

temperature range
Object mode: 

±

4% or 

±

2

(4

) whichever is greater

14

Содержание DET-1013b

Страница 1: ...e ear canal of human It converts the measured heat into a temperature reading displayed on the LCD When properly used it will quickly assess your temperature in an accurate manner The infrared thermom...

Страница 2: ...e clock setting section to learn how to setup the time in the first use The thermometer has been designed for practical use It s not meant to replace a visit to the doctor Please also remember to comp...

Страница 3: ...the unit Press Setting Button to select the unit you want After the unit is set press Mode Button the time format figure will appear Set the time format The device can display the time in either an AM...

Страница 4: ...ver on the probe firmly making sure the probe and the inner portion of the probe cover fit snugly 2 Press the The display is activated to show all segments Remarks a Probe cover sign stand for taking...

Страница 5: ...e Taking Hints 18 Temperature Taking Hints Figure 5 Figure 6 Bluetooth requirements The thermometer requires a bluetooth device with Bluetooth 4 0 or later Android 6 0 or later IOS 8 0 or later And wo...

Страница 6: ...mometer and probe cover in a dry location free from dust and contamination and away from direct sunlight 8 Periodic cleaning and disinfection of the device following use to prevent patient cross infec...

Страница 7: ...nctioning properly Read Illustration For Use thoroughly and take a new temperature measurement Replace two new alkaline batteries size AAA Allow the thermometer to rest in a room for at least 30 minut...

Страница 8: ...amage during shipping Include the original sales slip indicating the date of purchase a note describing the problem and your return address Send the thermometer prepaid and insured The lay operator or...

Страница 9: ...The device is an electrical medical product and is subject to special precautionary measures with regard to EMC which must be published in the instructions for use Portable and mobile HF communicatio...

Страница 10: ...c Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predic...

Отзывы: