Switching on 

Press “ ” key and the detector will be switched on after two beeps.

 

  Switching off

 

At measuring 

status, press and hold “ ” key down, the detector will be switched off 

after 5 seconds’ countdown; if the holding time for pressing “ ” key is not long enough, the 

detector will be brought back to the measuring status. 

  Low power 

Audible and visible alarm for undervoltage will be activated if the battery voltage is 

lower than the pre-set value. And the battery should be recharged at such situation. At 

undervoltage alarm, the screen will also display a detection failure symbol 

  besides the 

audible and visible alarm, and meanwhile the symbol 

  for detection pass will flicker 

together with the symbol 

  for battery undervoltage. When the power is exhausted, the 

symbol 

  for detection failure will flicker together with the symbol 

  for battery 

undervoltage, and 

the screen will display an indication of “battery off”, and then the 

detector will be switched off automatically. 

After an undervoltage alarm, the detector should be switched off and recharged for 

4~6 hours around. 

 

3.1 Functional description of the operating keys 

LCD Display 

 

 

The LCD screen has an 

auxiliary function of background 

light displaying which will be 

automatically activated at alarm 

when the gas concentration has 

exceeded the alarm point; at 

normal detecting in an 

environment with weak light, it 

can be activated by pressing UP 

Key or DOWN Key once. 

1. Detection pass   

 

2. Detection failure      3. 

Grade I Alarm 

4. Grade 

II Alarm             

5. STEL Value   

        6 .TWA Alarm 

7 Battery  

 

                8. %LEL (unit)                  9. %Vol (unit)

 

10. PPM (unit)

 

Содержание BX80

Страница 1: ...Portable O3 Ozone Tester Detector BX80 Operating instruction manual...

Страница 2: ...ontent 1 General 3 2 Technical performance and parameters 3 3 Operating instruction 4 4 Adjusting and calibrating 11 5 Daily use and maintenance 14 6 Storage and quality assurance 14 7 Complete set 15...

Страница 3: ...BX80 has a very clear and large LCD with audible and visible alarm and can detect accurate ozone level and send alarm to the operator in time for prevention even in very inimical work environment 2 T...

Страница 4: ...battery voltage TWA STEL Ex mark Ex ibdIICT4 Protection degree IP45 Work temperature 10 50 Work humidity 5 90 RH Size 126mm L x 66mm W x 32mm H Weight 220g with charger 3 Operating instruction Menu K...

Страница 5: ...her with the symbol for battery undervoltage When the power is exhausted the symbol for detection failure will flicker together with the symbol for battery undervoltage and the screen will display an...

Страница 6: ...cuit and the sensor s 2 Displaying the version of software and detection range 3 Displaying Grade I alarm point combustible gas noxious gas and high concentration O2 alarm and Grade II alarm point com...

Страница 7: ...ation of the gas and output a measuring result The detection principle of diffusion mode applies the above way to detecting the concentration of the surrounding gas with an advantage of swift response...

Страница 8: ...alarm symbol Grade I high concentration Grade II low concentration STEL and TWA will indicates as per the gas type and alarm grade If the concentration of the gas detected is beyond the measuring rang...

Страница 9: ...alarm indicating mode Mode conversion for alarm or non alarm Setting value for Grade I alarm point Setting value for Grade II alarm point Setting value for STEL alarm point Setting value for TWA alarm...

Страница 10: ...ration Alarm Indication at exceeding YES means the detector works with measuring and Vibration Alarm Indication the modification can be affirmed by or unchanged if not pressing Alarm Mode Select by pr...

Страница 11: ...te it usually The detector can be calibrated according to the instruction in this Section after it has been used for a long time or it is necessary to take a calibrating or readjusting Calibrating sho...

Страница 12: ...Zero point calibrating Sensitivity calibrating for O2 0 0 vol For noxious gas sensors the sensitivity calibrating should be carried out every 6 months The user should apply a standard gas to sensor c...

Страница 13: ...3 seconds the detector will be brought into sensitivity calibrating mode with audible indication The user can adjust the gas type for calibrating by UP Key or DOWN Key and affirm by Then it will have...

Страница 14: ...intenance or replace of any part should be carried out by a specially and well trained person with original spare parts from our company 5 5 Replace or recharge of battery should be carried out in a s...

Страница 15: ...environment will be beyond the guarantee scope 6 3 Our company will take no responsibility for the reliability of the detector if the user has carried out any repair or replace of any part beyond the...

Отзывы: