INFUSION PUMP
User Manual
Version: 3.1
Страница 1: ...INFUSION PUMP User Manual Version 3 1...
Страница 2: ...ing Packing list 9 Pump Installation 9 IV Set Installation 12 Chapter 4 Operation 13 Function key description 13 Operation Flowchart 14 Volume mode operation 15 VTBI Rate Time 15 Rate Time 16 Simple R...
Страница 3: ...ting 33 Chapter 9 Routine check 35 Chapter 10 Machine performance parameter General specifications 36 Response time of occlusion alarm 36 Chapter 11 Electrical specification 37 Chapter 12 Storage and...
Страница 4: ...heat them otherwise it will result in electrolyte leakage fire or explosion 10 Do not tear up the shell of batteries otherwise it will cause an explosion or chemical burn hazard 11 When plugging or un...
Страница 5: ...different tubing type and different drug liquid it may cause large deviations in infusion accuracy 26 if alarms are presented the related medical personnel should be take the related solutions in time...
Страница 6: ...Simple rate mode Rate 0 1 1200ml h VTBI 0 9999ml Time 1min 99h59min VOL 0 1 9999ml Drop mode Drops min Only For Option Choose mode needed VTBI time rate VTBI rate VTBI time Time rate Simple rate mode...
Страница 7: ...U Dual Lock Key By manual lock to avoid misuse Programmable Infusion can be programmed in 8 segments Library Built in 49 common drugs could up to 1000 kinds of drug History Record 1500 event records 2...
Страница 8: ...Exterior components names description Infusion tube parts description Handle LCD Key Fluid warmer Infusion set guide slot Bubble sensor Bottom pressure sensor Infusion set guide slot Up pressure Peris...
Страница 9: ...on infusion status icon pressure status icon IV Set brand specification and fluid warmer temperature 2 Infusion parameter Infusion pump structure Figure 4 Machine status bar Battery status Infusion se...
Страница 10: ...bling Packing list Please check the follow contents when unpacking 1 Infusion pump body 1 set 2 Drop sensor 1 set 3 Power cord 1 set 4 User s manual 1 copy Pump Installation Horizontal installation on...
Страница 11: ...10 Vertical installation on bracket Installation on cylinder...
Страница 12: ...nding plugs into the rear sockets of infusion pump according to the icon shown on pump back Notes if the drop sensor is not connected drop sensor failure will alarm 2 If not use the drop sensor can cl...
Страница 13: ...de then up occlusion happens and prompt alarm Purge the air inside the infusion tube make sure no bubble inside then close the adjustable clamp of ordinary PVC set to prevent loss of liquid Open the i...
Страница 14: ...s CFM Set Used to confirm and save the items chose In the menu press it to enter next level menu Press it in process of infusion can reset infusion rate Menu Exit On the condition of pause press it to...
Страница 15: ...r is normal etc after self check completed machine would enter infusion standby interface waiting for user to input infusion parameters Notes Not connect the AC power supply or the AC power is interru...
Страница 16: ...rface after volume mode is chosen VTBI Rate Time Move the cursor to corresponding input boxes of rate VTBI and time with the up or down key Set rate Set VTBI Set time if need Press Notes set every two...
Страница 17: ...I value would be displayed OFF Set rate Press Notes 1 Choose simple infusion mode should first move the cursor to VTBI input box with the up or down key long press for 1 seconds The VTBI value would b...
Страница 18: ...f 1ml h Infusion mode selection Press in starting up status display infusion mode interface Choose program mode with up or down keys Press to enter program mode interface VTBI Rate Setting Programming...
Страница 19: ...rogramming stage When no need more program infusion parameters press to enter the overview interface check whether the parameters of each programming stage are correct if some parameters of certain st...
Страница 20: ...onding input boxes with the up or down keys Dose unit selection When dose unit flashes press to choose the dose needed with the up and down key Input dose value press or the down key Input weight Pres...
Страница 21: ...Import Infusion mode selection Mode selection drug selection Press in starting up status display infusion mode interface Choose Drug Lib Mode with up or down key Press to enter the parameters input in...
Страница 22: ...drug needed press At concentration input box press then the concentration will displayed N A which means the unknown drug concentration Input dose rate then press Input weight if needed Press Choose...
Страница 23: ...process of infusion Set Bolus rate Set Bolus VTBI Press to begin Bolus Return to normal infusion after Bolus completed Manual Bolus Long press in process of infusion Machine would infuse according to...
Страница 24: ...management Infusion set selection Brand selection Choose infusion set type Exit to main interface Press Choose IV Set menu press Choose Brand menu press Choose IV Set brand needed press Choose IV Set...
Страница 25: ...re one electronic balance or measuring cup if choose electronic balance should put a vessel upon and reset the electronic balance before use 3 Close the adjustable clamp of infusion set waiting to be...
Страница 26: ...hether should calibrate again according to the actual value inputted Normally after 1 2 times calibration the accuracy would meet the requirement of deviation within 5 2 If alarm prompts in process of...
Страница 27: ...brand of tube type actually used is not in the brand list of machine user can add infusion set brand by tube type add operation The operation as follows Choose menu Input brand Press confirm to finis...
Страница 28: ...brand need be deleted Prompt delete or not Prompt deletion completed Choose Delete brand menu press Choose the IV Set brand need to be removed by using up and down keys press Choose according to the p...
Страница 29: ...xpel the bubbles inside infusion tube the infusion needle cannot be connected to patient when execute purge operation Mute Can suspend the alarm voice by using the Mute key when alarm prompts The mute...
Страница 30: ...menu of fluid warmer press Choose the menu of heating range press Press to return to infusion interface Pre alarm setting of infusion completion Press during infusion Choose the menu of infusion compl...
Страница 31: ...appears but alarm doesn t prompt please choose the level of small number When bubble alarm prompt in high frequency please choose the level of larger number Pre alarm setting of infusion completed Pr...
Страница 32: ...31 Time setting Set system time of machine Can input year month date hour minute second TFT setting Brightness adjustment of LCD Use up and down keys to adjust the brightness...
Страница 33: ...ory records in machine Choose menu Choose item Input date Check results Check results Press Choose record menu press Choose the item need to be checked press Input the date need to be checked press Ch...
Страница 34: ...d Check whether the infusion bottle bag is empty Check whether Murphy s dropper is misty or too tilted liquid flows down along the wall rather than dripping Up Occlusion High Yes Infusion inlet is not...
Страница 35: ...heck the infusion line is installed correctly Check whether the drip rate sensor is subjected to strong interference shaking Check whether the door pressing plate is loose Fluid Warmer Failure High Ye...
Страница 36: ...n set then boot Open pump door door open alarms and no liquid drops Close pump door and press clear key door open alarm disappears and no liquid drops Before boot Drop sensor Correctly install infusio...
Страница 37: ...unt Unit 0 1 9999 units Concentration Unit 0 01 1 unit Up pressure 100 300mmHg High median and low 3 options Down pressure 200mmHg 1200mmHg 21 options Bubble detection 50ul 800ul Pre alarm 5min 30min...
Страница 38: ...time 12 h Communication Can communication wired or wireless communication By RS232 Certificate of Compliance Comply with 93 42 CE medical directive IPX4 Degree of Protection Water splash proof Degree...
Страница 39: ...Direct sunlight or strong light 2 Direct contact with blowing hot wet air from heating electric stove or humidifier 3 Chemical storage place or noxious gas place 4 Place of Water seepage splash dusty...
Страница 40: ...acturer s Statement of Anti electromagnetic Interference apply to all the equipments and systems Infusion Pump is intended to be used in the following specified electromagnetic environments Buyers or...
Страница 41: ...ning equipment and systems Infusion Pump is intended to be used in the following specified electromagnetic environments Buyers or users should ensure that the pump is used in such environments Immunit...
Страница 42: ...ion distance m based on communication equipment frequency 150KHz to 80MHz 80MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 As to the...
Страница 43: ...inimum bias in infusion pump The testing programs for these results have been included in EN60601 2 24 IV set Drag Heart 20 drops ml standard PVC set The following graphs and curves show the overall c...
Страница 44: ...f compliance The equipments connected with the pump should comply with their respective IEC or ISO standards i e information processing equipment IEC60950 and the whole equipment should comply with th...
Страница 45: ...44 Product and service information Product name Infusion Pump Product standard EN IEC 60601 1 60601 2 24 EN IEC 60601 1 2 60601 1 8 Distributor By...