PEDs can record up to 9999 tasks. The four most recent tasks can be reviewed through the task details screen.
Attention
: All task data logging periods may be displayed when data is viewed in
DoseVision™.
Task details can be cleared from PEDs after data has been transferred
to DoseVision™.
2.7
Discrete mode
The discrete mode disables the audible and visual warnings of alarm conditions. For example, to avoid situations
which may cause distress to others or could draw unwanted attention. In the discrete mode, all the LEDs are
permanently switched off and the sounder is de-activated so that audible alarms are never activated. Also, the screen
is set to a dimmer setting. However, the vibration alert still activates and on-screen indications are still shown.
Attention
: If the screen-saver feature is switched on when using discrete mode, the screen wakes on an alarm in a
dimmed state.
The discrete mode must be enabled using
DoseVision™. Once enabled, the discrete mode icon is shown in the main
menu. When the discrete mode is switched on, a cross is added to the icon.
2.8
Handheld mode (PED+ and PED-ER+)
The handheld mode allows the PED+ and PED-ER+ to be used as a dose rate meters or survey meter. The PED is
effectively measuring H*(10) dose rate and therefore cannot be considered to be measuring personal dose.
Therefore, personal dose is not accumulated while in the handheld mode.
Attention: Personal dose is not accumulated while in handheld mode
In the handheld mode, the PED+ and PED-ER+ measures the dose rate which is compensated for
the fact that it is no longer being worn on the body. Therefore, the PED+ and PED-ER+ should be
held away from the body whenever possible.
The PED+ and PED-ER+ automatically flip the screen when switching to the handheld mode to encourage the user
to hold the PED+ or PED-ER+ with the GM-tube window facing forwards or towards the radiation area.