After placing the suture using standard continuous or 
interrupted technique load an additional clip and pass 
the loaded LAPRA-TY Suture Applier through the trocar.

Position the clip by sliding it down the suture until it makes 
contact with the tissue. 

NOTE:

 Excessive tension may 

cause clip slippage.

Close the applier jaws by squeezing the handle of the 
applier until the clip is visibly locked. Surgeon should 
verify closure and security.

LAPRA-TY® Suture Applier

Optimized Device Performance Guide

Grasp the applier by the handle and insert the jaws of 
the instrument into the individual cartridge slot. Ensure 
the tips of the applier are perpendicular to the surface 
of the cartridge.

Place a LAPRA-TY® Suture Clip approximately 5mm  
from the cut end of the suture. 

NOTE:

 Avoid placing 

suture near the latch in order to ensure latch security.

LAPRA-TY Suture Clips are not intended for use in 
clinical situations that require the use of suture material 
other than sizes 2-0, 3-0 or 4-0 Coated VICRYL® 
(polyglactin 910) Suture. Failure to use indicated suture 
type and size may result in insufficient clip security.