48    DePuy Synthes  User Manual  Small Battery Drive II

Troubleshooting

Any serious incident that has occurred in relation to the 
device should be reported to the manufacturer and   
the competent authority of the Member State in which 
the user and/or patient is established. 

Problem

Possible causes

Solution

Handpiece does not start up.

Battery is dead.

Charge the battery or replace it with  
a charged battery.

The tool was not cooled off after 
 sterilization.

Let the tool cool to room temperature.

Mode selector switch is on OFF.

Turn the mode selector switch to  
ON or  .

No contact between the handpiece  
and the battery pack.

Reinsert the battery pack or replace it.

Handpiece lacks power. 

Battery is dead.

Charge the battery or replace it with a 
charged battery.

Machine stops suddenly.

The machine has overheated  
(over-heating protection is activated).

Wait until the machine has cooled down.

Battery is dead.

Charge the battery or replace it with a 
charged battery.

Attachments cannot be 
 coupled to the unit.

The attachment coupling is blocked  
by deposits.

Remove solid objects with a pair  
of tweezers.

Warning:

 Immediately turn mode 

switch to locked position (off position). 
Remove solid particles with pickups. 
Check the release sleeve; clean and 
lubricate if necessary (Depuy Synthes 
Special Oil 519.97). Send machine to 
Depuy Synthes Service Center if 
necessary.

Difficulty removing 
attachments from drive unit

Coupling is blocked by residue

Warning:

 Immediately turn mode 

switch to locked position (off  
position). Remove solid particles with 
pickups. Clean and  lubricate coupling 
sleeve according to care and 
maintenance guidelines.

Drive unit coupling sleeve is damaged

Send damaged drive unit to Depuy 
Synthes Service Center

Tool (saw blade, drill, burr etc.) 
can not be coupled or only 
with difficulty.

Shaft geometry of the attachment or
tool is damaged.

Replace the attachment or tool or send
it to your Depuy Synthes service office.

Oscillating saw attachment 
 vibrates too much.

The saw blade locking mechanism  
is not tight.

Tighten the locking knob of the saw 
blade coupling.

The mode selector switch is set to  .

Turn the mode selector switch to ON.

Содержание DePuy Synthes Small Battery Drive II

Страница 1: ...Battery Driven Power Tool System for Small Bone and Sports Medicine Surgeries Small Battery Drive II User Manual...

Страница 2: ......

Страница 3: ...eam Attachments 19 Other Rotating Attachments 21 Saw Attachments 26 Care and Maintenance General Information 30 Cleaning and Disinfection 31 Preparation Prior to Reprocessing 31 Manual Cleaning Instru...

Страница 4: ...2 DePuy Synthes User Manual Small Battery Drive II Table of Contents System Specifications 50 Electromagnetic Compatibility 56 Ordering Information 60...

Страница 5: ...he Small Battery Drive II is intended to be used by qualified health care professionals e g surgeons operating room staff and individuals involved in preparation of the device Indications The Small Ba...

Страница 6: ...is cleaned disinfected and serviced after each use in accordance with the process defined in the Care and Maintenance section Compliance with these specifications can considerably extend the service...

Страница 7: ...ditions please refer to the section System Specification on page 51 Servicing Warning This system requires regular maintenance service at least once a year in order to maintain its functionality This...

Страница 8: ...protection rating according to IEC 60529 Explanation of the general symbols used The following symbols are applied to the device or individual components Information on additional symbols is given in...

Страница 9: ...Manufacturing date Non sterile Non sterile Sterilized using radiation Do not use if package is damaged Reference number Lot number Serial number Expiration date Packaging unit ISO 7000 2794 2009 02 Me...

Страница 10: ...e of the handpiece OFF or ON position Protective systems The Small Battery Drive II is equipped with three protective systems A thermal overload safety system that shuts off the tool if it becomes too...

Страница 11: ...working with the Small Battery Drive II system Only place the tool in an upright position when changing attachments or cutting tools intraoperatively The handpiece must be laid on its side when not i...

Страница 12: ...Li Ion battery 532 103 follow the guidelines provided in the STERRAD V PRO sterilization guide To ensure sterility of the battery casing the battery is inserted into the battery casing by two people o...

Страница 13: ...ons when inserting the unsterile battery into the sterile battery casing The circulating person must avoid all contact with the outside of the battery casing in order not to contaminate it Should the...

Страница 14: ...attery casing with the battery should be kept ready to ensure fast intraoperative change of batteries under sterile conditions Inserting the battery pack into the power tool Guide the battery pack bat...

Страница 15: ...72 hours Charging storing and using of Small Battery Drive II batteries 532 103 Charging Warning Only use the Depuy Synthes Universal Battery Charger II 05 001 204 to charge the battery Using a charge...

Страница 16: ...otate clockwise as usual Simultaneously pressing the top and bottom triggers causes the tool to immediately switch to oscillating mode The clamped tool oscillates clockwise anticlockwise The speed can...

Страница 17: ...n page 7 simultaneously and remove the attachment from the coupling Warnings To avoid injuries the locking mechanism of the tool has to be activated before every manipulation and before placing the to...

Страница 18: ...ed with Key clamping range up to B 7 3mm with reverse motion 350 rpm Red Other Rotating Attachments 532 011 Mini Quick Coupling 3500 rpm None 532 012 J Latch Coupling 3500 rpm None 532 022 Quick Coupl...

Страница 19: ...0 5 4 0mm 310 93K For drilling 05 001 253 0 5 7 3mm 510 191 For drilling Attachments Drill Attachments Inserting cutting tools Open the jaws of the Chuck using the appropriate key or by hand Insert th...

Страница 20: ...be taken when inserting screws with the drive unit Never fully insert screws with the drive unit The last turns or locking should always be done manually Always use an appropriate torque limiting att...

Страница 21: ...nto the coupling Maneuver the unlocking ring on the attachment backward and insert the tool such as a drill bit while rotating it slightly until it locks into place Release the ring Check if the tool...

Страница 22: ...Use the Chuck with Reverse motion 05 001 254 only with tools that are approved for such use Otherwise the tool may break with subsequent damage To ensure secure fixation of the instrument make sure th...

Страница 23: ...usting sleeve K Slightly press the adjusting sleeve axially against the handpiece and rotate the sleeve 2 Apply a slight amount of pressure to insert the Kirschner Wire from the front into the cannula...

Страница 24: ...to the Small Battery Drive II and the adapter to the Quick Coupling Position the Radiolucent Drive as far as it will go over the Quick Coupling and the adapter and rotate it into the desired working p...

Страница 25: ...e After the incision position the Radiolucent Drive and center the drill bit tip over the locking hole On the monitor of the image intensifier you can see both the drill bit and the target rings of th...

Страница 26: ...s can cause an overload Correcting the drilling angle when the cutting edges of the drill bit are completely in the bone Hitting the nail with the drill bit Drilling can be continued after having made...

Страница 27: ...th burrs of size M the burr is correctly clamped when the marking M on the burr shank is no longer visible Information on handling burrs Depuy Synthes recommends using a new sterile burr for each oper...

Страница 28: ...ade from the direction of the machine Changing the saw blade Fig 2 1 Lock the machine 2 Pull the locking knob O down and turn it counterclockwise 3 Lift and remove the saw blade 4 Use a slight amount...

Страница 29: ...tch in the bone Optimal saw performance is achieved by moving the power tool slightly back and forth in the plane of the saw blade so that the blade oscillates beyond the bone on both sides Very preci...

Страница 30: ...ade Pull the saw blade coupling toward the handpiece and insert the saw blade while rotating it slightly until it locks into the saw attachment coupling Release the saw blade coupling and check that t...

Страница 31: ...switch of the handpiece is in the OFF position and that the locking sleeve on the saw attachment is set to the unlock position Insert the saw attachment in any position into the attachment coupling of...

Страница 32: ...ials and personnel in the processing unit Any deviation by the processor from the instructions provided should be properly evaluated for effectiveness and potential adverse consequences Power tool uni...

Страница 33: ...nufacturer of the disinfectant Warnging Inspect battery for cracks and damage Precautions Do not use solvents to disinfect the batteries Battery poles must not contact water or solvents danger of shor...

Страница 34: ...er single use items or if reusable be decontaminated at least daily using a solution as detailed in section 3 Spray and wipe Brushes shall be inspected before daily use and discarded if they have degr...

Страница 35: ...trate Enzymatic Cleaner 1 8 oz per gallon and clean the device manually under running warm water for a minimum of 5 minutes Manipulate all moving parts under running water Use a soft bristled brush an...

Страница 36: ...urified water for a minimum of 2 minutes 10 Dry Dry device using a soft lint free cloth or medical grade compressed air If smaller devices or cannula tions contain residual water blow dry with medical...

Страница 37: ...attachments the cleaning brush 519 400 shown below should be used Notes Brushes and other cleaning tools shall be either single use items or if reusable be decontaminated at least daily using a solut...

Страница 38: ...2X Concentrate Enzymatic Cleaner 1 8 oz per gallon and clean the device manually under running warm water for a minimum of 5 minutes Manipulate all moving parts under running water Use a soft bristled...

Страница 39: ...be used for sterilization but is not required for machine washing Warning Do not wash the system in the Depuy Synthes Graphic Cases 06 550 060 06 550 200 and 06 550 210 Dimensions of the Washing Bask...

Страница 40: ...C Thermal disinfection 5 93 C Dry 40 90 C 10 Inspect the device Remove all the devices from the washing basket Inspect the cannulations coupling sleeves etc for visible soil If necessary repeat the ma...

Страница 41: ...tion for Use of the Depuy Synthes Special Oil 519 97 and the Small Battery Drive II Care and Maintenance Chart Lubricating the handpiece Fig 1 and 2 Lubricate the trigger shafts and then press the tri...

Страница 42: ...I Lubricating the battery casing Fig 3 and 4 Place oil on the seal of the cover and then evenly distribute the oil on the seal Lubricate the lock hinge and knob then actuate it several times Figure 3...

Страница 43: ...Coupling 05 001 250 05 001 251 Quick Coupling for DHS DCS Triple Reamers 532 015 Quick Coupling for Medullary Reaming 532 017 532 018 532 019 532 020 Lubricate the unlocking ring Move it back and fort...

Страница 44: ...does not require lubrication The power tools and attachments must only be lubricated with Depuy Synthes Special Oil 519 97 No oil from other manufacturer must be used Lubricants with other compositio...

Страница 45: ...remain blocked in the handpiece when pressing on them Check that no residuals prevent the movable parts from moving smoothly Check the coupling sleeves of the handpiece and attachments for smooth ope...

Страница 46: ...aturated steam forced air removal pre vacuum minimum 3 pulses Minimum 4 minutes Minimum 132 C Maximum 138 C 20 60 minutes Minimum 3 minutes Minimum 134 C Maximum 138 C 20 60 minutes Drying times gener...

Страница 47: ...inted on the packaging label Packaged and sterilized products should be stored in a dry clean environment protected from direct sunlight pests and extremes of temperature and humidity Use products in...

Страница 48: ...for damage resulting from improper use neglected or unauthorized maintenance Faulty devices must not be used If it is no longer possible or feasible to repair the tool it should be disposed of refer...

Страница 49: ...risk of infection in case of disposal Always discharge the batteries and isolate the contacts before disposal Risk of fire explosion and burns Do not disassemble crush heat above 60 C 140 F or incine...

Страница 50: ...s cannot be coupled to the unit The attachment coupling is blocked by deposits Remove solid objects with a pair of tweezers Warning Immediately turn mode switch to locked position off position Remove...

Страница 51: ...ce guidelines The triggers are difficult to move The trigger shafts need to be lubricated Lubricate the trigger shafts as described on page 39 It is difficult to couple the battery casing to the machi...

Страница 52: ...A C22 2 NO 60601 1 14 Medical electrical equipment Part 1 2 Collateral Standard Electromagnetic disturbances Requirements and tests IEC 60601 1 2 2014 Ed 4 0 EN 60601 1 2 2015 Medical electrical equip...

Страница 53: ...pressure 1060 hPa 1 06 bar 500 hPa 0 5 bar 1060 hPa 1 06 bar 500 hPa 0 5 bar Altitude 0 5000 m 0 5000 m Transportation Temperature Duration Humidity 29 C 20 F 72h uncontrolled 38 C 100 F 72h 85 60 C...

Страница 54: ...2 Continuously adjustable speed 0 3500 rpm Weight w battery and battery casing 925 g Operating voltage 14 4 VDC Battery capacity 1 2 Ah Battery type Li Ion Cannulation B 3 2mm Empty battery charging t...

Страница 55: ...needs to be taken into consideration during the planning of the surgical intervention Always check the temperature of the system to prevent overheating and possibly harming the patient or user Warning...

Страница 56: ...ry 2011 Precaution Carefully observe the above recommended maximum daily exposure time Handpiece Attachment Tool Sound Pressure Level LpA in dB A Sound Power Level LwA in dB A Max daily exposure time...

Страница 57: ...1711 5323 03 10 date of testing 18 February 2011 Handpiece Attachment Tool Vibration emission m s2 Max daily exposure Small Battery Drive II 532 110 2 5 8 h AO ASIF Quick Coupling 05 001 250 2 5 8 h...

Страница 58: ...sure that it is used in such an environment Emission test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The Small Battery Drive II system uses RF energy only for its in...

Страница 59: ...umidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines Not applicable Mains power quality should be that of a typical commercial or hospital environmen...

Страница 60: ...P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from xed RF trans...

Страница 61: ...distance according to frequency of transmitter m 150 kHz to 80 MHz d 0 35 P 80 MHz to 800 MHz d 0 35 P 800 MHz to 6 2 GHz d 0 7 P 0 01 4mm 4cm 7cm 0 1 11cm 11cm 22cm 1 35cm 35cm 70cm 10 1 11m 1 11m 2...

Страница 62: ...532 110 and 532 110 05 001 187 Burr Attachment for Nos 532 001 532 010 532 110 532 110 and 05 001 175 532 021 Oscillating Saw Attachment for Nos 532 001 532 010 532 110 532 110 and 05 001 175 532 023...

Страница 63: ...he brochure Working with the Radiolucent Drive Materials Device s Material s Standard s Drive units Stainless Steel ISO 7153 1 Aluminum DIN EN 573 PEEK n a Charger Battery Stainless Steel ISO 7153 1 a...

Страница 64: ...d with the devices identified in this brochure for additional information CAUTION Federal Law restricts these devices to sale by or on the order of a physician Some devices listed in this brochure may...

Отзывы: