IV Pole Setup
Z370584 Rev. D
4-4
Surgery Start Up
IV Pole Setup
The IV pole moves to the appropriate height automatically based on the settings of the
selected surgical mode. Follow these steps to set up the IV pole.
1.
Remove the protector from the tip of the drip chamber spike. With the bottle upright,
insert the drip chamber spike straight into the center of the rubber bottle stopper without
twisting.
2. Hang the bottle of balanced salt solution on the IV pole.
3. Squeeze the drip chamber to fill the drip chamber with fluid to the half-full level.
4. Raise or lower the pole if needed. Use the IV pole up and down arrows on the upper right
of the touchscreen. You can also use the rocker switch on the console.
5. Attach the I/A tubing to the handpiece.
NOTE: To ensure the bottle height does not change when you select I/A after vitrectomy
during surgery, check the Bottle Height Hold checkbox on the Vitrectomy Settings
screen.
Handpiece Setup
This section presents information about setting up the handpieces.
•
“Phacoemulsification Handpiece Setup – VERITAS Swivel” on page 4-5
•
“Phacoemulsification Handpiece Setup – WHITESTAR and ELLIPS FX” on page 4-7
•
“Irrigation/Aspiration Handpiece Setup” on page 4-9
•
“Diathermy Handpiece Setup” on page 4-9
•
“Vitrectomy Cutter Setup” on page 4-10
IMPORTANT:
Before you insert the spike into the bottle, shake the irrigation drip chamber at the end of the
irrigation tubing to confirm that the irrigation valve moves. If the valve does not rattle, the
valve cannot operate properly and irrigation cannot flow.
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...