Warnings, Cautions, and Other Residual Risks
Z370584 Rev. D
Warnings
2-3
•
Never handle the foot pedal by its cable.
•
Do not attempt to carry the foot pedal by the heel when the heel insert is attached. The
heel insert could become loose causing the foot pedal to drop.
•
Do not place the foot pedal on a wet surface.
•
Securely attach plugs, handpiece cables, and other connectors.
•
Do not have any other Bluetooth
®
devices in the same area as the remote and the console
(other remote controls, foot pedals, cell phones, or headsets, for example) as the pairing
operation will fail. The software can detect a maximum of nine (9) devices.
•
Never attempt to straighten a bent phaco ultrasonic tip. This might produce a broken tip
when you apply ultrasound.
•
Perform functional checks of the system before first use of the day.
•
If you do not hear a tone when you press the foot pedal and volume adjustment is
unsuccessful, the mode is not functioning properly. Refer to
Troubleshooting, and Diagnostics"
.
•
Only use the JJSV specified replacement battery in the Advanced Linear Pedal to ensure
correct system performance. See
“Charging Options for Wireless Foot Pedals” on page 10-4
Warnings
•
All personnel who might operate this equipment must read and understand the
instructions in this manual before they use the system. Failure to do so might result in the
improper operation of the system. Only a trained licensed physician can use this device.
•
Do not modify the VERITAS
™
Vision System.
•
The system comes equipped with a 3-prong power plug which you must plug into an outlet
with a ground receptacle. If the plug does not fit the outlet, contact an electrician.
•
To avoid the risk of electric shock, this equipment must only be connected to a supply
mains with protective earth.
•
Ensure the balanced salt solution bottle is at or above the eye level of the patient.
•
The surgical staff must monitor the balanced salt solution bottle height and fluid level at all
times. A low bottle or empty bottle affects the fluid balance and the intraocular pressure
(IOP) while aspirating. Low bottle height or low or empty bottle fluid level can result in:
Inadvertent chamber shallowing or collapse
Aspiration or abrasion of the iris or other tissue
An ultrasonic wound heating commonly called wound burn (extreme case)
•
Do not attempt to use the system if the system fails to perform properly as stated in this
manual.
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...