Fuse Replacement Procedure
Z370584 Rev. D
10-2
Error Messages, Troubleshooting, and Diagnostics
•
If there is no phacoemulsification, confirm there is no damage to the phaco tip/handpiece
by dropping or misuse.
•
If no irrigation occurs, shake the drip chamber to confirm that the ball or the valve moves
freely. If there is no rattle sound, replace the drip chamber with another disposable tubing
pack.
Fuse Replacement Procedure
If the system does not turn on when you press the power switch, you have confirmed that
you attached the power cord properly, and you plugged the cord in, check to see if the fuse
is bad.
Table 10.1 – Fuse Specifications
To replace the console fuses:
1.
Unplug the system electrical power.
2. Unplug the power cord from the back panel.
3. Locate the fuse holder on the rear panel of the system (see
).
4. Use a small screwdriver to gently pry open the cover and expose the fuse holder.
5. Gently pry out the fuse holder.
6. Remove the bad fuse and replace the fuse with a new fuse (see value and size in
).
7. Replace the fuse holder. Ensure the arrows point to the right side of the back panel. Tilt
the fuse holder slightly to the right and push in.
8. Push the fuse holder cover up and in until the cover snaps closed.
9. Reattach the power cord to the back panel.
10. Plug the system into an electrical receptacle.
WARNING!
To prevent the risk of fire or damage to the instrument, replace the fuses with the
exact type and rating as indicated below (check the voltage sticker on the back panel
of the system to confirm your system voltage).
Voltage
Quantity
Fuse Specifications
100/120/240
2
6.3A, 250V, Bussmann GDA
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...