Cleaning and Maintenance
Important:
Clean the accessories before using the nebulizer for the first time or after each use
following these instructions: wash all the pieces, except for the air filter, with warm water
and detergent. Rinse with warm tap water to remove the detergent.
Dip the pieces into a clean bowl containing 3 parts warm water and 1 part vinegar, let
soak for 30 minutes. Disinfectant can also be used.
Wash your hands and remove the pieces from the cleaning solution. Rinse them with
warm tap water and let them air dry. Keep the pieces into a tightly sealed bag. Do not
dry them with a towel. Clean the outer surface with a dry piece of cloth.
WARNING: In order to avoid infections, it is highly recommended to clean the pieces
after each use. Only nebulize solutions prescribed by a doctor and follow the doctor’s
instructions carefully. Do not use balm substances
Air Filter Change:
1. The air filter must to be replaced if it turns dirty or torn.
2. Remove the air chamber from the medication cup cover.
3. Take out the filter and place a new one.
4. Put all the pieces back .