10 | 

P a g e

 

DRAFT 

Power frequency 

(50/60Hz) Magnetic 

field 

IEC 61000-4-8 

30 A/m 

30 A/m 

If  display  is  distorted,  it  may  be 
necessary  to  position  the  Model 
Myolift QT. Further from sources of 
power frequency magnetic fields or 
to  install  magnetic  shielding.  The 
power  frequency  magnetic  field 
should be measured in the intended 
installation location to assure that it 
is sufficiently low. 

NOTE: UT is the a.c mains voltage prior to application of the test level. 

 

Recommended  Separation  Distances  Between  Portable  and  Mobile  RF 
Communications Equipment and the Myolift QT System

 

The  Myolift  QT  device  is  intended  for  use  in  an  electromagnetic  environment  in  which  radiated  RF 
disturbances  are  controlled.  The  user  of  the  Myolift  QT  device  can  help  prevent  electromagnetic 
interference  by  maintaining  a  minimum  distance  between  portable  and  mobile  RF  communications 
equipment (transmitters) and the Myolift QT  device as recommended below, according to the maximum 
output power of the communications equipment.

 

 

Rated maximum 

output power (W) of 
transmitter

 

Separation distance (m) according to frequency of transmitter

 

150 kHz to 80 MHz

 

d = 1.17 √P

 

80 MHz to 800 MHz

 

d = 1.17 √P

 

800 MHz to 2.7 GHz

 

d= 2.33 √P

 

0.01

 

0.12

 

0.12

 

0.23

 

0.1

 

0.37

 

0.37

 

0.74

 

1

 

1.17

 

1.17

 

2.33

 

10

 

3.70

 

3.70

 

7.37

 

100

 

11.70

 

11.70

 

23.30

 

For transmitters rated at a maximum output power not listed above, the recommended separation 
distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the 
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to 
the transmitter manufacturer.

 

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

 

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is a

ected by 

absorption 

and reflection from structures, objects, and

 people.

 

 

 

 

Содержание Myolift QT

Страница 1: ...Instruction Manual Myolift QT...

Страница 2: ...ndicatorwillexplainpossible Safetyinfractionsthat couldhavethe potentialtocauseminortomoderateinjuryordamagetoequipment WARNING Textwitha WARNING indicatorwillexplainpossibleSafetyinfractionsthatwill...

Страница 3: ...IONS 8 8 3 GUIDANCE AND MANUFACTURIN S DECLARATION 8 9 CONTROLS FUNCTIONS 11 10 How to Operate Device 12 10 1 DEVICE ON 12 10 2 HOW TO ESTABLISH CREDENTIAL WITH APP 12 10 3 HOW TO CONNECT THE DEVICE W...

Страница 4: ...nstruction Manual 3 P a g e DRAFT 16 WARRANTY 19 16 1 WARRANTY COVERAGE 19 16 2 WARRANTY DISCLAIMERS 20 16 3 LEGAL REMEDIES 20 16 4 WARRANTY PERFORMANCE 20 16 5 CUSTOMER SERVICE 20 16 6 OUT OF WARRANT...

Страница 5: ...ng or not operating properly Do not disassemble the device as this may cause damage malfunction electrical shock fire or injury There are no user serviceable parts inside the device 1 2 CAUTIONS Not f...

Страница 6: ...eneath the applicator may be caused by use of other powered muscle stimulators If you experience skin irritation or burning discontinue the use of the device immediately If you experience a persistent...

Страница 7: ...th the type of users iPhone Android The User has to download the application via Google Play store Android App Store iOS and connect the device via Bluetooth of their phone and make their profiles acc...

Страница 8: ...dicates that the device is charging up And it will show the intensity Low or high Note At the time of treatment steady green light will be shown on device 6 INTENDED USE The intended use of the device...

Страница 9: ...ment observe and verify normal operation of the system in the configuration in which it will be used prior to using it Hand held and mobile RF communications equipment may a ect the normal function of...

Страница 10: ...ontact 15 kV air 8 kV Contact 15 kV air Floors should be wood concrete or ceramic tile If floors are covered width synthetic material the relative humidity should be at least 30 Electric fast transien...

Страница 11: ...e between portable and mobile RF communications equipment transmitters and the Myolift QT device as recommended below according to the maximum output power of the communications equipment Rated maximu...

Страница 12: ...to comply with the limits for a Class B digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential in...

Страница 13: ...ere is a one touch sensor in the middle of the device where the user has to long press 4 to 5 seconds it until the beep sound will not come to ON OFF the device 10 2 HOW TO ESTABLISH CREDENTIAL WITH A...

Страница 14: ...tructions a Write your first name b Write your last name c Your mail ID and password d Select the type of skin e Select the focus areas as per your need Once you completed with all the above steps you...

Страница 15: ...ce Once you open the application below screen will come Note Make sure that your device is visible to nearby device s 10 4 DEVICE OPERATE RUN TIME Once you ON the device it will run in its pre defined...

Страница 16: ...in in a precise way to stimulate the different muscles requiring treatment 12 DEVICE OPERATION SETTING Note 1 When the user will open the application on his phone there is one video that will help the...

Страница 17: ...e DRAFT 12 2 FOR TREATING FACE LIFT Tap on learn option and select the intensity low or high through application Once you select the intensity level then tap on start the treatment option Treatment t...

Страница 18: ...ead mask Conductive gloves and Metal balls as per his need 13 DEVICE TREATMENT Facial muscles create chronic movement in the face and are used minute by minute Facial muscles begin losing elasticity d...

Страница 19: ...ANING The device has to be always turned off by means of the mains switch when cleaning The mains power switch OFF The Device must be cleaned thoroughly prior to usage to remove visible soil To clean...

Страница 20: ...Always wash the selected treatment area with mild soap and water and dry it thoroughly before applying lip and eye mask and after removing them Do not stretch the skin or lip and eye mask while apply...

Страница 21: ...roduct is returned to the distributor The repaired or replacement product will be in warranty for the balance of the one year warranty period and an additional one month period No charge will be made...

Страница 22: ...re designated by the MANUFACTURER as repairable by SERVICE PERSONNEL Manufactured By JOHARI DIGITAL HEALTHCARE LTD G 582 584 EPIP BORANADA JODHPUR 342012 RAJ INDIA TOLL FREE 1800 102 8684 Email info j...

Страница 23: ......

Отзывы: