7.4.0 Specifications
7.4.1
Performance
Accuracy +/-
0.015
inches
Repeatability 0.001%
of
full
span
Linearity
0.020%
of
full
span
Rate
of
Change
(Max)
6
inches
per
second
Refresh
Rate 10
times
per
second
Initiation
0.00
seconds
Damping 0.00
to
1.00
@
0.01
second
increments
1.00
to
25.0
@
1.00
second
increments
Unusable
Region
1.00
inch
(at
end
of
probe)
Dead
Zone
0.00
inch
(user
specify)
Humidity
0.00
–
99.0%
(non-condensing)
7.4.2
Electrical
Input
12-30
VDC
(24
VDC
Nominal)
Output 4-20
mA
Resistance
600
Ohms
(max)
@
24
VDC
Power Consumption 0.66 Watt (30 VDC x 0.022 ohms = 0.66 W)
Error
Signal 3.60
or
22
mA
Interface
3
button
keypad,
HART,
or
PACT
software
Display 2
line
8
character
LCD
Values Inches
or
centimeters,
percent
of
level,
mA
7.4.3 Ratings
Process
Pressure 300
PSIG
@
100°
F
LGT
only
Ambient Temperature -40° to +160° F (-40° to +70° C) Electronics
-10°
to
+160°
F
(-40°
to
+70°
C)
LCD
Readout
Process Temperature
Standard -40°
to
+235°
F
(-40°
to
+114°
C) All
Models
HT Version -40° to +400° F (-40° to +214° C) ILT & MGT
7.4.4 Enclosure
Type
Single
Compartment
Material
Cast
Aluminum
(optional
316
SS)
Finish Polyester
Powder
Coat
Rating FM,
CSA
Approved,
Type
Nema
4X
7.4.5 Sensor
Material
316
SS,
(optional
CPVC,
Hastelloy,
Alloy
20)
Length 6.00
to
300
inches